background image

Gilson Company, Inc. 

GilSonic AutoSiever: GA-6

Page 9

6.1  Setting Maximum Amplitude
A. Using the REC/

 and SAVE/

 arrow keys, adjust 

  the amplitude level to an initial low number, such as 5, and 

press the Run/Stop button to start.  

 CAUTION: The Electroformed cloth in Precision Sieves is 

especially delicate and may be destroyed if the amplitude 

setting is too high. While it is impossible to state an exact 

maximum value, an amplitude setting of 25 or more for 

sieves of 20µm or less should be approached incrementally. 

Coarser Electroformed sieves will be more robust, as will 

most wire-woven sieves, but it is still possible to damage 

them with excessively high settings or prolonged run-times. 

If you are unable to achieve satisfactory separations in 15 

minutes or less at amplitude settings under 50, please 

consult a Technical Support representative at Gilson. 

Attempting higher settings may void the warranty.

B.  Gradually increase the amplitude until sample material on 

the top sieve flows smoothly in a gentle fluid action. Note 

the final number on the display for later programming as the 

maximum amplitude value.

6.2  Establishing Phase Times
Phase times require experimentation to find the best values, 

but the settings below are generally a good starting point: 

 

 

  Time “A” (Ramp-up) 0.5 minutes

 

  Time “B” (Time at Max. Amplitude) 5.0 minutes

 

  Time “C” (Cleanup) 0.3 minutes

C.  Press MAN/AUTO to return the unit to automatic operation. 

D.  Press PGM until the Time A indicator LED is blinking. 

E.  Use the keypad to enter the time for the amplitude to ramp 

up from 0 to the established maximum setting. A default 

value of 0.1 is already programmed. 

  NOTE: Time is displayed in 0.1 minute (6 second) 

increments. 1 minute, 18 seconds would be displayed as 

1.3. Digits entered appear in the right-hand space of the 

display and “push” other digits to the left. When entering 

new values, it may be necessary to enter 0’s until the display 

is cleared.

F.  Press PGM to accept the displayed value. The setting will be 

stored and the Max. Amplitude LED will begin to blink. 

G.  Enter the maximum amplitude setting as determined in 

manual mode above and press PGM. The value will be stored 

and the Time B LED will start blinking. 

H.  Continue entering and storing values for Time B and Time 

C phases. After a value for Time C is entered and stored, the 

total time for the test (A+B+C) will be displayed. 

I.  Select one of the three Tapping Action buttons to program 

tapping during operation.

  NOTE: For most materials, operation with both horizontal 

and vertical tapping (

/

) insures the best separation. 

Some problem materials may respond best using only 

vertical tapping or none at all.

J.  To execute the entered program one time, simply press the 

Run/STOP button. The AutoSiever will run the sequence and 

stop automatically at completion. 

  NOTE: After running through a test cycle at these initial 

settings, the individual fractions can be weighed, then 

tested again for an additional minute at the previously 

determined amplitude setting. If the amount passing a 

given sieve increases less than about 1%, then the original 

settings are adequate. If there is more than an additional 

1% passing, the process of increasing test times in one-

minute increments should be continued. If the material 

being tested is exceptionally dense or lightweight, 

prone to agglomerate or gather static charges, time 

and amplitude settings may have to be increased. (See 

Caution note in Section 6.1 A.)

K.  To store the entered program into memory, press SAVE and 

enter a number from 0 to 9 for the memory location. 

Summary of Contents for GA-6

Page 1: ...ever GA 6 OPERATING MANUAL Rev 01 2019 PHONE 800 444 1508 P O Box 200 Lewis Center Ohio 43035 0200 740 548 7298 E mail customerservice gilsonco com Product Web Page www globalgilson com FAX 800 255 53...

Page 2: ...Test Phase Indicator Lights 6 4 4 Operating Mode Indicator Lights 6 5 0 Preparing the Sieve Stack Assembly 7 6 0 Establishing Test Sequences 8 6 1 Setting Maximum Amplitude 9 6 2 Establishing Phase Ti...

Page 3: ...from No 20 850 m to No 635 20 m or precision electroform nickel mesh 150 m to 5 m Electroformedmeshsizesof150 mandlargerareavailable for 2 m precision when needed Largerwire wovensievesfromNo 3 1 2toN...

Page 4: ...S unplug or disconnect machine from the power source when the unit is not in operation 2 0 SAFETY INSTRUCTIONS 3 0 UNCRATING SETUP Afterinspectingtheshipmentforexternaldamage immediately unpack the Gi...

Page 5: ...advance to the next entry This function is also used to review saved values in an existing program REC AND SAVE 4 0 CONTROL DISPLAY FUNCTIONS Fuse Voltage Selector Block TheRecallandSavefunctionsareus...

Page 6: ...ethatdecreaseswithparticlesize Thegradualdecrease attheendofthetestservestocleanupseparationofnear sized particles on each sieve Max Amplitude ThisLEDisonatanytimetheprogrammedmaximumamplitude setting...

Page 7: ...ede airflow enough to damage sieves B On a balance sensitive to at least 0 01 gram determine and record the tare mass of parts that will contact the sample during testing Diaphragm Top Cone Each Sieve...

Page 8: ...maximum F PlacetheColumnLockovertheDiaphragmandpressdown until the Stack Locking Bars snap onto the top ring of the Fines Collector Holder G Insert the Sieve Stack Assembly into the sieving chamber o...

Page 9: ...n 0 1 minute 6 second increments 1 minute 18 seconds would be displayed as 1 3 Digits entered appear in the right hand space of the display and push other digits to the left When entering newvalues it...

Page 10: ...C through J or select a previously stored program from one of the ten memory locations by pressing REC and the memory location C Press the RUN STOP button to initiate the test The AutoSieverwillexecu...

Page 11: ...ower cord from the rear of the unit B Remove the 7 Phillips head screws on the top of the unit and remove the case top and handle C Noteorientationofthespeakerandremovethetwolugged wires from the term...

Page 12: ...70 181 GAA 71 65 181 GAA 72 63 181 GAA 72A 60 181 GAA 73 55 250 GAA 74 53 250 GAA 74A 50 250 GAA 75 45 250 GAA 76 40 300 GAA 77 38 300 GAA 77A 35 300 GAA 78 32 300 GAA 78A 30 300 GAA 79 25 400 GAA 80...

Page 13: ...nic AutoSiever GA 6 Page 13 Column Lock GAA 85 Diaphragm GAA 3 Top Cone GAA 87 Standard Spacer GAA 88 Fines Collector GAA 4 Fines Collector Holder GAA 89 Complete Sieve Stack Assembly GAA 2 12 0 REPLA...

Reviews: