PIPETMAN® M
|
USER’S GUIDE
Chapter 17
TROUBLESHOOTING
In case of malfunction, first reset the pipette by pressing on the push button and the
mode button simultaneously for at least eight seconds.
If the problem persists, you may consult the following table that identifies potential
problems and their solutions.
WARNING
Before returning any pipette to your local Gilson service center, ensure that
it is completely free of chemical, biological, or radioactive contamination.
Please use the included safety bag.
PROBLEM
POSSIBLE CAUSE
PAGE
Pipette is leaking
sample
Worn O-ring
Pipette won’t
aspirate
Unscrewed lower part
of tip holder
Damaged or corroded piston
Damaged tip holder
Improper repair or assembly
Connecting nut is loose.
Software needs to be reset
Noisy operation
Piston needs lubricating
Pipette is
inaccurate
Improper repair or assembly
Unscrewed lower part
of tip holder
Pipette is out of adjustment
Connecting nut is loose
PROBLEM
POSSIBLE CAUSE
PAGE
Pipette is not precise
Unscrewed lower part of tip holder
Incorrect operator technique
Worn O-ring
Connecting nut is loose
Damaged or corroded piston
Damaged tip holder
Tips fall or don’t fit
Low quality tips
Damaged tip holder
Damaged tip ejector
Ejector spacer is damaged
Tip ejector is loose
Dirty tip holder
No OLED display
Pipette is in sleep mode
Battery needs recharging
Software needs to be reset.
No operation possible
Battery needs recharging.
Software needs to be reset
Calibration impossible
Software needs to be reset
Chapter 18
SPECIFICATIONS
PIPETMAN M is a high quality pipette that offers excellent accuracy and precision; it is
fully compliant with ISO 8655 and is CE marked.
The figures given in the
PIPETMAN M Maximum Permissible Errors
Table 2 on page 30
were obtained using “PIPETMAN DIAMOND Tips”. These figures are only guaranteed
when using genuine PIPETMAN DIAMOND Tips.
Each pipette is inspected and validated by qualified technicians according to the Gilson
quality system. Gilson declares that its manufactured pipettes comply with the
requirements of the ISO 8655 standard, by type testing.