Operation
MetPak Base Station User Manual
©
Gill Instruments Ltd
Last Change
Page 44
22 April, 2013
3.4.3
Monitoring MetPak Base Station Output Data
Using HyperTerminal.
NOTE.
. This section describes the procedure using Windows HyperTerminal.
Alternatively Gill Wind Software (
http://www.gill.co.uk/main/software.html
) can be
used as a Terminal Emulator (note Tools feature not applicable). Select the
required COM port and click on OK, the required 19200 Baud Rate is already
selected (or use the Baudrate drop down) then continue as from step 7.
Establishing a HyperTerminal Connection
To monitor MetPak Base Station data using HyperTerminal:
1.
Select an available Com port and connect the MetPak Base Station unit.
2.
Run
Hypertrm.exe (
Typically accessed via:-
Start | All Programs | Accessories | Communications | Hyperterminal).
3.
Select
New Connection
from the
File
menu.
4.
Enter a name for the connection, for example
MetPak Base Station
and,
optionally, select an icon to represent it. Click on the
OK
button.
5.
In the
Connect To
dialog box, change
Connect Using
to
COMX
(where
X
is
the number of the selected COM port). Click on the
OK
button.
6.
In the
COMX Properties
dialog box, adjust the
Port
settings
to match the
MetPak Base Station unit’s settings. The default settings are :
Bits per second
19200
Data bits
8
Parity
None
Stop bits
1
Flow Control
None
7.
Click on the
OK
button.
Message Reporting
MetPak Base Station has two reporting modes:
Continuous mode
In continuous mode (the default) MetPak Base Station reports continuously. So, if
you have set up the connection correctly, you should now see data strings scrolling
across the HyperTerminal window. For example:
2 Axis wind Sensor.
Q,270,000.01,1013.8,042.7,+022.0,+008.8,+11.6,00, 7E
3 Axis Wind Sensor
Q,270,000.01,-
000.02,1013.8,042.7,+022.0,+008.8,+11.6,00, 63
Polled mode
If the unit has been set to polled mode type ?Q (where Q is the unit node letter) to
prompt one line of the data string. Note, allow 9.5 seconds after switching on the
unit before sending a poll command.
When polled, allow 570 milli-seconds for the unit to respond with a line of data. If
the MetPak Base Station detects a pressure/temperature/humidity sensor fault
then allow 4 seconds for an error data output.