Giga-tronics GT-1050A Microwave Power Amplifier
Operation
Operation Manual, Part Number 34817, Rev. C, March 2011
9
Chapter 3
Operation
3.1
Operating Safety and Instructions
DO NOT EXCEED AN INPUT LEVEL OF +20 dBm INTO THE GT-1050A. EXCEEDING
THIS LEVEL CAN DAMAGE THE GT-1050A MICROWAVE POWER AMPLIFIER.
WHEN ENERGIZED, THE GT-1050A IS CAPABLE OF SUPPLYING POWER THAT
CAN CAUSE DAMAGE OR INJURY. TAKE THE FOLLOWING PRECAUTIONS TO
ENSURE SAFE SETUP AND OPERATION:
Verify that all cables, connectors, and equipment connected to the
GT-1050A are in good condition.
Do not make connections to equipment while the output of any item of
equipment is energized.
Table 4: Operate the GT-1050A
Operate the GT-1050A
Step
Action
1.
Verify that the POWER switch on the rear of the unit is OFF.
Plug the included external power supply into a source of AC power that meets the
specifications for power in Table 10 on page 12.
Plug the DC cable from the external power supply into the con9 V on the rear of the
unit.
Put the POWER switch on the GT-1050A in the ON position.
NOTE:
For best results, let the GT-1050A warm up for one minute after switching the AC
power ON.
2.
Verify that the output of the microwave signal source is NOT energized before continuing to
the next step.
3.
Connect the equipment to the GT-1050A according to your application.
NOTE:
Verify all mating connectors are 50 Ohm, and that they are in good condition.
4.
Energize the output of the microwave signal source.
5.
Adjust the output of the microwave signal source until the output from the GT-1050A is at the
desired level.
End of Procedure
CAUTION