Remote Operation
Publication 31470-001, Rev C, November 2, 2017
3-137
Table 3-21: Measurement Error Code Returned in Position AA
Error Code Message
Notes
00
No Error
All OK
01
Cannot zero sensor A
Ensure no RF power to sensor A
02
Cannot zero sensor B
Ensure no RF power to sensor B
03
Sensor A not connected to Calibrator
Connect sensor A to Calibrator
04
Sensor B not connected to Calibrator
Connect sensor B to Calibrator
05
Cannot Cal Sensor A
Check sensor A connection to Calibrator;
reference must be 1.00 mW
06
Cannot Cal sensor B
Check sensor B connection to Calibrator;
reference must be 1.00 mW
21
Over limit
An over-limit condition has occurred
23
Under limit
An under-limit condition has occurred
26
Sensor A unable to synchronize Burst Average Power
(BAP) measurements to a pulse stream
Check measurement setup and RF signal
27
Sensor B unable to synchronize Burst Average Power
(BAP) measurements to a pulse stream
Check measurement setup and RF signal
31
No sensor on Channel A
Connect sensor A, or change channels if B is connected
32
No sensor on Channel B
Connect sensor B, or change channels if A is connected