Power Sensors
Manual 20791, Rev. C, June 2001
A-3
See NOTES below
NOTES:
1.
The K connector is electrically and mechanically compatible with the APC-3.5 and SMA connectors. Use a Type N(m) to
SMA(f) adapter (P/N 29835) for calibration of power sensors with Type K(m) connectors.
2.
For frequencies above 8 GHz, add power linearity to system linearity.
3.
Square root of sum of the individual uncertainties squared (RSS).
4.
Cal Factor numbers allow for 3% repeatability when connecting attenuator to sensor, and 3% for attenuator measurement
uncertainty and mismatch of sensor/pad combination.
Table A-2: Power Sensor Cal Uncertainties
Freq. (GHz)
Root Sum of Squares (RSS) Uncertainties (%)
3
Lower
Upper
80301A
80302A
80303A
80304A
80310A
80313A
80314A
80320A
80323A
80324A
80321A
4
80330A
80333A
80334A
Min
1
1.04
1.64
1.58
1.58
4.54
1.58
1
2
1.20
1.73
1.73
1.73
4.67
1.73
2
4
1.33
1.93
1.91
1.91
4.89
1.90
3
6
1.41
2.03
2.02
2.01
5.01
2.01
6
8
1.52
2.08
2.07
2.06
5.12
2.06
8
12.4
1.92
2.55
2.54
2.53
5.56
2.53
12.4
18
2.11
2.83
2.80
2.79
5.89
2.78
18
26.5
——
3.63
3.68
3.62
——
3.59
26.5
40
——
6.05
5.54
5.39
——
5.30
Summary of Contents for 8003
Page 3: ......
Page 14: ...8003 Precision Scalar Analyzer x Manual 20791 Rev C June 2001 ...
Page 16: ...8003 Precision Scalar Analyzer xii Manual 20791 Rev C June 2001 ...
Page 18: ...8003 Precision Scalar Analyzer xiv Manual 20791 Rev C June 2001 ...
Page 20: ...8003 Precision Scalar Analyzer xvi Manual 20791 Rev C June 2001 ...
Page 36: ...8003 Precision Scalar Analyzer 1 16 Manual 20791 Rev C June 2001 ...
Page 142: ...8003 Precision Scalar Analyzer 2 106 Manual 20791 Rev C June 2001 ...
Page 232: ...8003 Precision Scalar Analyzer 4 18 Manual 20791 Rev C June 2001 ...
Page 252: ...8003 Precision Scalar Analyzer B 16 Manual 20791 Rev C June 2001 ...
Page 258: ...8003 Precision Scalar Analyzers Index 6 Manual 20791 Rev C June 2001 ...