![GHM GREISINGER Operating Manual Download Page 31](http://html1.mh-extra.com/html/ghm/greisinger/greisinger_operating-manual_2220427031.webp)
G 1501
Error and system messages | 8
B-H86.0.12.DB2-2.1
31 / 37
ERR.1
Measuring range
exceeded
Measurement too
high
Incorrect electrode
connected
Electrode or product
defect
Stay within allowable meas-
urement range
Check electrode
Send in for repair
ERR.2
Measuring range is
undercut
Measurement too
low
Incorrect electrode
connected
Electrode or product
defect
Stay within allowable meas-
urement range
Check electrode
Send in for repair
SYS ERR
System error
Error in the product Switch product on/off
Replace batteries
Send in for repair