G 1501
Operation | 7
B-H86.0.12.DB2-2.1
29 / 37
Displayed value = measured value
–
T.OF
Gradient correction °C:
Display = (measured value
–
T.OF
) * (1 +
T.SL
/ 100)
Gradient correction °F:
Display = (measured value
– 32 °F –
T.OF
) * (1 +
T.SL
/ 100) + 32 °F
–
Zero point correction
T.OF
to
0.00
–
Gradient correction
T.SL
to
0.00
–
Display unit
UNIT
to °C
–
Display in ice water -0.2 °C
–
Display in ice water setpoint
T.OF
= 0.0 °C
–
Display in water bath 36.6 °C
–
Display in water bath setpoint
T.SL
= 37.0 °C
–
T.OF
= display zero point correction
– setpoint zero point
–
T.OF
= -0.2 °C
– 0.0 °C = -0.2 °C
–
T.SL
= (setpoint gradient correction / (display gradient correction
–
T.OF
)
– 1) *100
–
T.SL
= (37.0 °C / (36.6 °C
– (-0.2)) -1) *100 = 0.54
The changed value is saved and the
Configuration
menu is closed.
NOTE
If the product is switched off without saving the configuration, the last save value is re-
produced on the next start-up of the product.
Example calculation
Outcome of an action