E36.0.01.6C-01
Installation and
Operating Manual GIA 0420 … - ex
page 16 of 20
Ex protection:
II 2 G Ex ia IIC T4
EC-Type Examination Certificate: BVS 11 ATEX 1 333 X
Connection data:
Umax:
28 V
Imax:
100 mA
Pmax:
1.2 W
Max. effective internal capacitance: Ci = 13 nF
additional for the switching output: Ci = 4.5 nF
Maximum effective internal inductance of the device are negligibly small
When making connections to the switching output the connections must be within the same
intrinsically safe circuit like the measuring signal
!
11 Decommissioning, reshipment and disposal
11.1 Reshipment
All devices returned to the manufacturer have to be free of any residual of measuring media and other
hazardous substances. Measuring residuals at housing or probe may be a risk for persons or
environment.
Use an adequate transport package for reshipment, especially for fully functional devices. Please
make sure that the device is protected in the package by enough packing materials.
11.2 Disposal instructions
The device must not be disposed in the unsorted municipal waste!
Send the device directly to us (sufficiently stamped), if it should be disposed.
We will dispose the device appropriate and environmentally sound.