4 | Operation
G 1610
12 / 26
B-H86.0.2X.DK2-3.0
4.2
Adjustment of the temperature measuring input
The temperature input can be adjusted with the zero point correction and the gradient
correction. If an adjustment is made, you change the pre-adjusted factory settings.
This is signalled with the display text
T.OF
or
T.SL
when switching on.
1.
Switch the device off.
2.
Hold the
down button
and press the
On/Off button
briefly to switch on the device
and open the
Adjustment
menu.
3.
The display shows the first parameter. Release the
down button
.
Parameter
Values
Meaning
_
T.OF
Zero point correction of the temperature
0.00
No zero point correction
-5.00
…
5.00
Zero point correction in °C. (at °F -9.00 .. 9.00)
_
T.SL
Gradient correction of the temperature
0.00
No gradient correction
-5.00
…
5.00
Gradient correction in %
Formula used by device:
Temperature = °C: Display = (measured value -
T.OF
) * (1 +
T.SL
/ 100)
Temperature = °F: Display = (meas. value - 32 °F -
T.OF
) * (1 +
T.SL
/ 100) + 32 °F