4 | Operation
G 1610
10 / 26
B-H86.0.2X.DK2-3.0
_
4
Operation
4.1
Opening the configuration menu
1.
Press the
Function key
for 2 seconds to open the
Configuration
menu.
2.
(ONF
appears in the display. Release the
Function key
.
_
Parameter
Values
Meaning
__
INP
Measuring unit
SAT
%
Oxygen saturation in percent
(ONC
mg/l
Oxygen concentration in mg/l
(ONC
ppm
Oxygen concentration in ppm
__
SET
.
P
Pressure
500 .. 4000
Environmental pressure in hPa, corresponding to
mbar
__
SAL
Salinity correction
0
..
70
Salinity in the measuring medium in PSU, corre-
sponding to g/kg
POFF
Shut-off time
OFF
No automatic shut-off
15
,
30
,
60
,
120
,
240
Automatic shut-off after a selected time in minutes,
during which no buttons have been pressed