HD2305.0
- 8 -
V1.6
change the compensation temperature using the arrows (
and
). Confirm using
DA-
TA/ENTER
. The display stops blinking, and the temperature displayed is used for compensa-
tion. If the temperature probe is not present, to change the unit of measurement between °C and
°F, it is necessary to press
twice
the
°C/°F
key.
MENU:
to access to the menu press simultaneously
DATA+°C/°F
: the
Probe Type
will be
displayed.
use the
and
arrows (respectively located above the HOLD and REL keys) to
modify
the displayed value;
press
DATA/ENTER
to
confirm
the modification and go onto the next item;
press
CLR/ESC
to
cancel
the modification;
to
exit
the menu, press the
°C/°F
key again.
For more info, please see following chapter.
Press this key to start the calibration procedure (vd. par. pH).
The "
/ REL-mV
" key is used for the following functions:
REL-mV:
it displays the difference between the current value and that measured on pressing
the key. The "REL" message is displayed on the left. To return to the normal measurement,
press the key again. It only works with
mV
, so to enable the REL function select the "mV" unit
of measurement using the "
pH-mV
" key.
:
Once the MENU has been opened with the °C/°F/MENU key, the
key will allow to de-
crease the value of the selected parameter. During electrode calibration, it allows to decrease the
nominal buffer value
/ REL-Mv key
CAL key