![GHM DeltaOHM HD2305.0 Operating Manual Download Page 19](http://html1.mh-extra.com/html/ghm/deltaohm-hd2305-0/deltaohm-hd2305-0_operating-manual_2220416019.webp)
HD2305.0
- 19 -
V1.6
INSTRUMENT
SIGNALS
AND
FAULTS
The following table lists all error indications and information displayed by the instrument and sup-
plied to the user in different operating situations:
Display indications
Explenation
>>>_PRBE_TYPE
type of probe connected
BATT TOO LOW - CHNG NOW
battery discharged - replace it immediately
BAUDRATE >>>
baud rate value
BUFR_1
first buffer value
BUFR_2
second buffer value
BUFR_3
third buffer value
CAL
LOST
Program error: it appears after turning on for a few seconds. Contact the instru-
ment's supplier.
CAL
blinking
Calibration not completed correctly.
CAL ERR
calibration error
ERR
This message appears if the pH measurement exceeds the -2.000pH…19.999pH
limits, if the mV measurement exceeds the
2.4V limits, if the temperature
probe, already detected by the instrument, is disconnected. At the same time an
intermittent beep is issued.
FUNC CLRD
max, min and average values cleared
NEW_PROB_DET
new probe detected
OFFS
offset
OVER
Measurement overflow: this appears when the probe measures a value exceed-
ing the measurement range or the mV measurement is included in the
+2.0…+2.4V range.
PLS_EXIT >>> FUNC
RES_FOR_FACT ONLY
please exit using ESC >>> function reserved to factory calibration
PRBE_SER #### ####
serial number #### #### of the connected probe
PROB
ERR
A probe with SICRAM module has been inserted when not admissible for that
specific instrument.
PROB COMM LOST
This appears if the probe, has already been detected by the instrument, but is
disconnected. At the same time an intermittent beep is issued.
SLPE%
gain %
SYS
ERR
#
Instrument management program error. Contact the instrument's supplier and
communicate the numeric code # reported by the display.
UNDR
Measurement overflow: this appears when the mV measurement is included in
the -2.4…-2.0V range.