GHM Delta OHM PMBsense-A Operating Manual Download Page 9

 

PMsense 

-   9   -

 

V1.0

 

5

 

PROPRIETARY PROTOCOL 

To use the proprietary protocol, it is necessary to connect the instrument to the PC via 
a  RS485/USB  (e.g.  RS51K)  or  RS485/RS232  converter  and  use  a  standard  serial 
communication  program.  In  the  serial  communication  program,  set  the  COM  port 
number  to  which  the  instrument  is  connected  and  the  communication  parameters  as 
follows: 

 

If  the  MODBUS-RTU  protocol  is  set  as  the  operating  protocol  in  the  instrument 
(default), set the Baud Rate 57600 and the parameters 8N2 in the serial commu-
nication  program,  then  power  cycle  the  instrument  and  send  the  command 

@

 

within 10 seconds from the instrument power on. 

 

If the proprietary protocol is already set as the operating protocol in the instru-
ment,  it  is  possible  to  operate  with  Baud  Rate  57600  and  parameters  8N2  by 
sending  the  command 

@

  within  10  seconds  from  the  instrument  power  on,  or 

you  can  let  the  10  seconds  pass  without  sending  the  command 

@

  and  operate 

with the communication parameters set in the instrument (default 19200, 8E1). 

To  change  the  instrument  configuration,  the  serial  command 

CAL  USER  ON

  must  be 

sent first (the instrument replies with USER CAL MODE ON). The command CAL USER ON 
is automatically disabled after a few minutes of inactivity. If the settings should be only 
read, the command CAL USER ON is not required. 

Below is the list of the serial commands. 

Instrument information: 

Command 

Reply

 

Description 

 

G0 

Model 

Instrument model 

G1 

&

Revision

Instrument hardware revision 

G2 

SN=nnnnnnnn 

Instrument serial number 

G3 

Firm.Ver.=x.y 

Instrument firmware revision 

G4 

Firm.Date=yyyy/mm/dd 

Date of firmware revision 

GC 

Fact.Calib.Date= yyyy/mm/dd 
User.Calib.Date= yyyy/mm/dd 
Cal.Mode=

Factory

 or 

User

 

Date of factory calibration 
Date of user calibration (CO

2

 only) 

Type of calibration active (CO

2

 only) 

Protocol:

 

Command 

Reply

 

Description 

 

&| 

Keeps the proprietary protocol operational even after 10 
seconds from instrument power on. It must be sent within 
10 seconds from instrument power on. 

DPn 

&| 

Sets the operating protocol: 

 

Proprietary if n=0 

 

MODBUS-RTU if n=1 

Default 

: MODBUS-RTU (n=1) 

RP 

& n| 

Reads the operating protocol set in the instrument. 

SM 

&| 

Activates the MODBUS-RTU protocol immediately. 

CMAn 

&| 

Sets the instrument address for the MODBUS-RTU pro-
tocol to n. 
The address should range within 1 and 247. 

Default 

: 1 

RMA 

& n| 

Reads the instrument address for the MODBUS-RTU pro-
tocol. 

Summary of Contents for Delta OHM PMBsense-A

Page 1: ...Operating manual PM CO2 Transmitter PMsense www deltaohm com English Keep for future reference Companies Brands of GHM...

Page 2: ...TENTS 1 INTRODUCTION 3 2 TECHNICAL SPECIFICATIONS 4 3 INSTALLATION 5 3 1 ELECTRICAL CONNECTIONS 5 4 CONFIGURATION AND MEASUREMENT 8 5 PROPRIETARY PROTOCOL 9 6 MODBUS RTU PROTOCOL 15 7 SAFETY INSTRUCTI...

Page 3: ...II proprietary protocol A version with two additional 4 20 mA 0 10 V on request when ordering analog outputs is available The two analog outputs can be independently associated with any of the measure...

Page 4: ...ponse time 120 s air speed 2 m s Long term stability 5 of measurement 5 years Temperature drift 1 ppm C General specifications Output RS485 with Modbus RTU or ASCII proprietary protocol Only PMsense A...

Page 5: ...scription 3 1 ELECTRICAL CONNECTIONS The transmitter has a M12 8 pole circular connector and uses the CPM12 8DA op tional cables with M12 8 pole connector on one side and open wires on the other side...

Page 6: ...e short circuited inside the transmitter The cable shield is not connected to the M12 connector To ensure a good noise immunity it is recommended to connect the cable shield to ground GND RS485 connec...

Page 7: ...T1 GND Vdc AOUT2 Fig 3 4 current analog outputs connection Voltage analog output optional only PM B sense A By default the voltage analog outputs are 0 10 V with 0 V 0 ppm 10 V 1000 g m3 The analog ou...

Page 8: ...erval can be configured respectively by using the CPLS and CPLP commands of the proprietary protocol or the holding reg isters with address 15 and 16 of the MODBUS RTU protocol In the measurement mode...

Page 9: ...USER ON must be sent first the instrument replies with USER CAL MODE ON The command CAL USER ON is automatically disabled after a few minutes of inactivity If the settings should be only read the comm...

Page 10: ...ber of data bits is fixed to 8 Default 8E1 n 2 RMP n Reads the setting of parity and stop bits CMWn Sets waiting time after transmission with MODBUS RTU protocol Immediate reception if n 0 violates pr...

Page 11: ...nce between analog output 1 and associated physical quantity 4 mA 0 V Min quantity 20 mA 10 V Max quantity if n 0 20 mA 10 V Min quantity 4 mA 0 V Max quantity if n 1 Default Direct correspondence n 0...

Page 12: ...the analog output 2 CA2SOn Sets the direct or inverse correspondence between analog output 2 and associated physical quantity 4 mA 0 V Min quantity 20 mA 10 V Max quantity if n 0 20 mA 10 V Min quanti...

Page 13: ...the information obtainable with the RA2T RA2L e RA2H commands Reading of the measurement information Command Reply Description S0 Disable the sending of the measurement information S1 Enable the sendi...

Page 14: ...er must first be placed in an environment with a known CO2 concentration the value n entered in the command The value t in the reply to the command indicates the number of seconds required for the tra...

Page 15: ...16 bit Integer 2 1 PM2 5 in N of particles ml Measurement updated according the operating mode set with the CPLS CPLP and CPS commands 16 bit Integer 3 2 PM10 in N of particles ml Measurement updated...

Page 16: ...10 s Available in continuous measurement mode 16 bit Integer 19 18 PM1 0 in N of particles ml Measurement averaged over 15 min and updated every minute Available in continuous measurement mode 16 bit...

Page 17: ...waiting time after transmission with MODBUS RTU protocol 0 immediate reception default 1 waiting 3 5 characters Bit 4 3 Enable disable the offset of the analog output 1 0 offset disabled 0 20 mA or 0...

Page 18: ...hysical quantity associated to the analog output 1 If the output is associated to PM the value is expressed as a number of tenths e g 5 0 5 g m3 If the output is associated to CO2 the value is express...

Page 19: ...0 1000 0 g m3 of PM 16 bit Integer 16 15 Setting of the PM measurement mode 0 continuous 1 at cyclic intervals default 16 bit Integer 17 16 Setting of the cycle interval in seconds for the PM measurem...

Page 20: ...ell as the specific measures described in this manual are followed The instrument proper operation and operating safety can be ensured only in the cli matic conditions specified in this manual Do not...

Page 21: ...DA x ca ble has to be ordered separately PMBsense A PM1 0 PM2 5 PM10 and CO2 transmitter RS485 output with MODBUS RTU or proprietary protocol two 0 4 20 mA 0 10 V on request when ordering analog outpu...

Page 22: ...PMsense 22 V1 0 NOTES...

Page 23: ......

Page 24: ...parties is excluded from the guarantee The instrument shall be returned FREE OF SHIPMENT CHARGES to your dealer The jurisdiction of Padua applies in any dispute The electrical and electronic equipmen...

Page 25: ...e result of the constant development of the product This may produce some differences between the information written in this manual and the instrument you have purchased We cannot completely exclude...

Page 26: ...GHM GROUP Delta OHM Delta Ohm S r l a socio unico Via Marconi 5 35030 Caselle di Selvazzano Padova ITALY Phone 39 049 8977150 Fax 39 049 635596 www deltaohm com info deltaohm com V1 0 09 11 2020...

Reviews: