PMsense
- 13 -
V1.0
Command
Reply
Description
CA2Hn
&|
Sets n as the maximum value of the measuring range of
the physical quantity associated to the analog output 2.
If the output is associated to PM, the value must be ex-
pressed as a number of tenths (e.g. n=50 to indicate 5.0
µg/m
3
). If the output is associated to CO
2
, the value must
be expressed as a number of units (e.g. n=800 to indicate
800 ppm).
Default
: 10000 (=1000 µg/m
3
of PM)
RA2H
& n|
Reads the maximum value of the measuring range of the
physical quantity associated to the analog output 2.
RA2F
&
Quantity
Minimum value
Maximum value
Simultaneously provides the information obtainable with
the RA2T, RA2L e RA2H commands.
Reading of the measurement information:
Command
Reply
Description
S0
&|
Disable the sending of the measurement information.
S1
&|
Enable the sending of the measurement information
every second.
For the meaning of the information sent by the instru-
ment, see the command S5.
S5
&
Measurements
Prints the list of the measurement information sent by
the instrument, in the following sequence:
PM measurement error (0=no, 1=yes)
PM1.0 in µg/m
3
PM2.5 in µg/m
3
PM10 in µg/m
3
CO
2
in ppm
Atmospheric pressure in hPa (internal sensor for
CO
2
measurement compensation)
Field not used
Field not used
Field not used
Power supply voltage
Internal board temperature
Note: the PM measurement is updated according the oper-
ating mode set with the CPLS, CPLP and CPS commands.