![GHM Delta OHM LPPYRA13 Operating Manual Download Page 23](http://html1.mh-extra.com/html/ghm/delta-ohm-lppyra13/delta-ohm-lppyra13_operating-manual_2220413023.webp)
LPPYRA13
- 23 -
V2.3
The extended commands allow setting the temperature unit of measurement (if the
temperature sensor is present). To change the unit of measurement:
1)
Send the command aXSCAL USER ON! (note: a=instrument address).
2)
Send the command aXSCFD! (to set °C) or aXSCFE! (to set °F).
3)
Send the command aXSCAL END!
For more information about the SDI-12 protocol, visit the website "www.sdi-12.org".