![GHM Delta OHM HD52.3D Series Operating Manual Download Page 25](http://html1.mh-extra.com/html/ghm/delta-ohm-hd52-3d-series/delta-ohm-hd52-3d-series_operating-manual_2220406025.webp)
HD52.3D…
- 25 -
V2.2
Parameters for SDI-12 mode:
Command
Reply
Description
CU3Ac
&|
Sets the SDI-12 address to c value
The address is an alphanumeric character
ranging within 0…9, a…z, A…Z. Default : 0
RU3A
& c|
Reads the SDI-12 address set in the instru-
ment
General parameters:
Command
Reply
Description
CGHn &|
Enables/disables
heating:
Disables if n=0
Enables if n=1
Default : Enabled (n=1)
RGH
n|
Reads heating enabling state set in the in-
strument
CWCnnnn
&|
Sets wind speed threshold to nnnn value (in
hundredths of m/s)
Value should range within 0 and 100 hun-
dredths of m/s (= 0…1 m/s)
Default : 20 (= 0.2 m/s)
(see Note 3)
RWC
& nnnn|
Reads the wind speed threshold value set in
the instrument (in hundredths of m/s)
CWaLnnn
&|
Sets time interval for the calculation of aver-
age speed and average direction to nnn value
Value should range within 1 and 600 s.
If the value is greater than 10 s, it must be
an integer multiple of 10.
note:the value can be greater than or equal to 10
s starting from firmware version 2.20
Default : 1 s
RWaL
& nnn|
Reads the time interval for the calculation
of average speed and average direction set
in the instrument
CWaMn
&|
Sets the method for the calculation of
average speed and average direction:
If n=0: scalar mean. The average inten-
sity is calculated as average of intensities
without consideration of
direction. The
average direction is calculated as average
of directions, and expressed according to
the extended characteristic
(see Note 4)
.
If n=1: vector mean. The average of the
coordinates along each measurement axis
is calculated. The average intensity and
the average direction are those deter-
mined by the two average coordinates.
Default : vector mean (n=1)
Summary of Contents for Delta OHM HD52.3D Series