![GHM Delta OHM HD52.3D Series Operating Manual Download Page 23](http://html1.mh-extra.com/html/ghm/delta-ohm-hd52-3d-series/delta-ohm-hd52-3d-series_operating-manual_2220406023.webp)
HD52.3D…
- 23 -
V2.2
Parameters for NMEA mode:
Command
Reply
Description
CU4Bn
&|
Sets Baud Rate for NMEA mode to:
2400 if n=1
4800 if n=2
9600 if n=3
19200 if n=4
38400 if n=5
57600 if n=6
115200 if n=7
Default : 4800 (n=2)
RU4B
& n|
Reads Baud Rate setting for NMEA mode
CU4In
&|
Sets interface for NMEA mode to:
RS232 if n=0
RS485 if n=1
RS422 if n=2
Default : RS485 (n=1)
RU4I
& n|
Reads interface setting for NMEA mode
CU4Mn
&|
Sets parity and stop bits for NMEA mode
to:
8N1 if n=0 [No parity, 1 stop bit]
8N2 if n=1 [No parity, 2 stop bits]
8E1 if n=2 [Even parity, 1 stop bit]
8E2 if n=3 [Even parity, 2 stop bits]
8O1 if n=4 [Odd parity, 1 stop bit]
8O2 if n=5 [Odd parity, 2 stop bits]
The number of data bits is fixed to 8
Default : 8N1 (n=0)
RU4M
& n|
Reads current setting of parity and stop
bits for NMEA mode
CU4Rnnn
&|
Sets transmission interval of string with
measurements in NMEA mode to nnn sec-
onds
The interval should range within 1 and 255
seconds
Default : 1 second
RU4R
& nnn|
Reads setting of transmission interval of
string with measurements in NMEA mode