Ensure the welds are clean and perfect. Once the welding operation is complete, close the two
shells, insert the gasket in the module
,
screw the fairlead. At the other end of the module, enter the
ring with the O-Ring as indicated in the picture.
Make sure the cable is not twisted while you are screwing the fairlead. Now the probe is ready.
Direct connection of the 4-sensor to a DIN45326 connector
Sensor 4 wire Pt100 can be welded
directly to the DIN45326 female
connector’s pins without having to
resort to the TP47 board. The 4 wires
of Pt100 must be soldered as shown
in the diagram to the left.
To use this type of probe, set the
“Probe Type” item in the menu as
stated in page 8.
Probe Pt100 is acknowledged by the
device upon start-up: insert the probe
when the device is on and only start
up after insertion.
Sensor
Direct connection to the connector
Pt100 4 wires
Female Connector on the welding