The probes are already factory calibrated. The calibration data are stored in the SICRAM module
memory.
TP3575 and TP3276.2 globe thermometer probes
Screw the globe to probe stem.
HP3201.2 and HP3201 natural ventilation wet bulb probes
•
Remove the sensor cap (the cap is not screwed).
•
Insert the cotton wick, previously dipped with distilled water, into the temperature probe. The
cotton wick must protrude from the probe for about 20 mm.
•
Fill the reservoir up till ¾ with
distilled water
.
•
Replace the cap.
•
Warning
: keep the probe vertical to prevent water from leaking.
Note: over time the cotton wick tends to calcify (harden): it must be replaced periodically.
TP3204S natural ventilation wet bulb probe
•
Remove the sensor cap (the cap is not screwed).
•
Insert the cotton wick, previously dipped with distilled water, into the temperature probe. The
cotton wick must protrude from the probe for about 20 mm.
•
Replace the cap.
•
Fill the bottle with 500 cc of
distilled water
.
•
Turn the probe over and firmly screw the bottle to the probe reservoir.
•
Turn the probe quickly (to avoid water spillage).
•
Secure the probe to the
HD32.2.7.1
support by using the two screws at the bottom of the
probe.