HD2328
- 13 -
V1.7
INSTRUMENT
SIGNALS
AND
FAULTS
The following table lists all error indications and information displayed by the
instrument and supplied to the user in different operating situations:
Display indications
Explanation
1ST_MEAS UP DOWN
Correct the first point using the arrows
/
2ND_MEAS UP DOWN
Correct the second point using the arrows
/
BATT TOO LOW
CHNG NOW
Indication of insufficient battery charge appearing on
turning on. The instrument issues a long beep and turns
off. Replace the batteries.
BURN
The probe is disconnected or faulty.
CAL
LOST
Program error: it appears after turning on for a few
seconds. Contact the instrument's supplier.
CAL FACT
Factory calibration.
CAL USER
User calibration.
CAL_MODE
Calibration mode
ERR
This message appears when a wrong function is called:
for example, should the A-B key be pressed or the AVG
function be enabled when one of the two probes is not
connected.
FUNC CLRD
Max, min and average values clearing
OVER o UNDR
Measurement overflow: indicates that the probe is
measuring a value exceeding the measuring range.
PLS_EXIT >>> FUNC
RES_FOR_FACT ONLY
Please exit using ESC >>> function reserved to factory
calibration
SEL CHAN
Input channel selection
SEL MEAS 1/2
Select the first/second calibration point
SEAL MEAS UNIT
Selection of the unit of measurement
SEAL_TYPE_TC
Selection of the type of thermocouple
SYS
ERR
#
Instrument management program error. Contact the
instrument's supplier and communicate the numeric code
# reported by the display.
INSTRUMENT
STORAGE
Instrument storage conditions:
•
Temperature: -25…+65°C.
•
Humidity: less than 90%RH without condensation.
•
Do not store the instrument in places where:
•
humidity is high;
•
the instrument may be exposed to direct sunlight;
•
the instrument may be exposed to a source of high temperature;
•
the instrument may be exposed to strong vibrations;
•
the instrument may be exposed to steam, salt or any corrosive gas.
The instrument case is made of ABS plastic: do not use any incompatible solvent for
cleaning.
Summary of Contents for Delta OHM HD2328.0