To  ensure  flawless  usage  of  our  devices,  the  following  points  must  be  observed  during 
assembly and installation: 
 

4.1 

Mechanical assembly  

  Take  the  appropriate  measures  to  prevent  freezing  of  the  medium.  Should  the  device  later  be 

subjected  to  an  ambient  temperature  <4°C,  no  operation  with  pure  water  may  be  performed. 
Frost damage may be caused by water remaining in the device. 

  Note the permitted installation positions from the product information. 

  The device may never serve as a fixed point. 

  The connector flanges/threads must be compatible. 

  A filter should be provided for heavily spoiled media. 

  Take note of the stated operating pressure as well as the permissible temperature range. 

  Thermal expansion of the pipework must be balanced out by compensators. 

  The device can be aligned horizontally or vertically.  

  The direction of flow is indicated on the device by an arrow. The arrow points in the direction of 

flow. With the HV device the flow in both directions is possible. 

  When tightening the screw joint the connection must be locked so that no torque is transferred to 

the body.  

 

This equipment is designed for low-viscosity media. Highly viscous medium delays the 
start of rotor/turbine.

 

5.1 

Activation 

Before activation note the limit values stated in the product information, such as for pressure 
stage, temperature range or maximum permitted flow. 
 

  If present, the transport lock must be removed. 

  When operating with protective covers, remove them immediately before assembly.

 

  Rinse the pipe system before assembly to remove any spoiling. 

  Observe the direction of flow. This is indicated on the device with an arrow and must run parallel 

to the pipe axis after assembly. 

  Properly tighten the device during assembly. 

  Endurance must be assessed with aggressive media. 

  To  avoid  heavy  pressure  surges  we  recommend  that  you  fill  and  ventilate  the  system  before 

activation. Pressure surges may not exceed the value of PN/PS. 

  Ensure that assembly is conducted properly. 

                                            

1

 The device FH has no hatch/scaling. 

Structure of the devices

1

  

Assembly and installation 

Activation, operation and maintenance 

Connectio
n outfeed 
side 

 

 

Figure 1: Schematic diagram of the devices PO/WR1/FR/RM und HV 
 

PO/WR1/FR/RM

 

Body 

Rotor 

Sight glass 

 

HV 

Connection 

Connection 

Turbine 

Sight glass 
 

Connection 

Summary of Contents for PO

Page 1: ... 4 7 Accessories 4 8 Device transportation and storage 4 9 Return 4 10 Disposal 4 11 Disclaimer 4 This operating manual must be locally stored so that it can always be accessed All procedures described within this operating manual may only be performed by trained personnel authorised by the operator with corresponding protective gear 1 1 Safety signs and symbols Danger signs and symbols are marked...

Page 2: ...s with applications the necessity arises to ensure minimal care e g with lubrication or coolants With the flow devices PO WR1 FR RM and HV a visual flow monitor of fluid media can be facilitated The rotor turbine rotation visualizes the volume flow 2 1 Active principle Rotor The liquid medium enters into the housing and takes the visible arranged rotor in rotation The frequency of the rotor allows...

Page 3: ... is possible When tightening the screw joint the connection must be locked so that no torque is transferred to the body This equipment is designed for low viscosity media Highly viscous medium delays the start of rotor turbine 5 1 Activation Before activation note the limit values stated in the product information such as for pressure stage temperature range or maximum permitted flow If present th...

Page 4: ...ing regulation A return form can be downloaded on our homepage at http www ghm messtechnik de en downloads ghm forms html Acceptance from GHM Honsberg can only occur if 1 a filled out form is provided with each return 2 packing is applied that prevents damage to the device 3 the device is cleaned so that no dangers are present 4 the filled out form and if necessary hazardous materials a safety dat...

Reviews: