H69.0.01.6C-02 

 Operating Manual GMH 285 / GMH 285-BNC 

page 7 of 8 

 

 

_____________________________________________________ 

_____________________________________________________________________________ 

 

8  Adjustment 

The instrument can be adjusted, assuming that: Reliable references are available, such as ice-
water regulated precision water baths or similar. 

Displayed value °C =(measured value °C - OFFS) * (1 + slope correction / 100 )        
Displayed value °F =(measured value °F - 32 °F - OFFS) * (1 + slope correction / 100 )       

 

Follow these instructions to adjust the device: 

  Switch the device on. 

  Press 

 and 

 together until 

the first parameter “

0FFS

” is displayed.  

  Set parameter with 

 = up or 

= down  

  Jump to the next parameter by pressing 

Parameter  Value 

Information

 

Key 

 

Keys 

 

 

 

 

OFFSET correction     

factory setting: oFF = 0.0°C

 

OFF

 

No offset correction 

-5.0 

 5.0 

°C  

or 

-9.0

9.0 

°F

 

Value of offset correction 

Slope correction 

factory setting: oFF= 0%

 

OFF

 

No slope correction  

-5.00 

 5.00 

Value of slope correction in % 

Press 

 again to store changed settings, the device restarts (segment test). 

NOTE:

 

If there is no key pressed within the menu mode within 2 minutes, the configuration will 
be cancelled, the entered settings are lost! 

9  Error and System  Messages 

    ----

  

No temperature probe connected 
Or at GMH 285 BNC: cable/sensor broken 
Or at GMH 285:          cable/sensor shorted 

   ERR.  1

 

  Value exceeding measuring range, value too high or cable/sensor broken 

   ERR.  2

    Value exceeding measuring range, value too low or cable/sensor shorted 

   ERR.   7

    System error – the device has detected a system error (device defective or not within 

working temperature) 

  

 

The blinking bAt display indicates low battery voltage, device will continue to work for a 
short time. 

 

  

The battery is consumed and has to be changed. Measurements are no longer 
possible.  

10 Accuracy Inspection: Adjustment /Update Service 

You can send the device to the manufacture or retailer for adjustment and inspection.  
Moreover the manufacturer can do the latest software update. This ensures that future 
improvements are provided to owners of older devices in a cost-saving way. You can display the 
current software version if you do not release the on/off button after you switched the device on, but 
hold it for more than 5 seconds. (i.e. “r. 1.0”) 

Summary of Contents for GREISINGER GMH 285

Page 1: ...r 26 D 93128 Regenstauf 49 0 9402 9383 0 49 0 9402 9383 33 info greisinger de as of version V1 0 GMH 285 GMH 285 BNC Operating Manual Pt1000 Precision Thermometer For exchangeable probes with alarm WEEE Reg Nr DE 93889386 GMH 285 GMH 285 BNC GMH GREISINGER GMH GREISINGER ...

Page 2: ...VICE 7 11 DISPOSAL NOTES 8 12 TECHNICAL DATA 8 1 Designated Use The device is measuring temperature in C or F A suitable Pt1000 temperature probe has to be connected for operation The safety requirements see below have to be observed The device must be used only according to its intended purpose and under suitable conditions Use the device carefully and according to its technical data do not throw...

Page 3: ... safety may be a risk if there is visible damage to the device the device is not working as specified the device has been stored under unsuitable conditions for a longer time In case of doubt please return device to manufacturer for repair or maintenance 3 WARNING Do not use these products as safety or emergency stop devices or in any other application where failure of the product could result in ...

Page 4: ...table or attached to a belt etc The device with pop up clip at position 90 can be set up on a table etc The device with pop up clip at position 180 can be suspended from a screw or the magnetic holder GMH 1300 Device attached to a belt Device set up on a table Device suspended from magnetic holder GMH 1300 5 Start of Operation Switch the device on with the key After segment test the device display...

Page 5: ...idered Exceeding the ranges at least causes a wrong measurement it may even damage the probe permanently Often it also has to be considered that the temperature range is just valid for the probe tube plastic handles can t stand the same high temperatures Therefore the tube length should be selected long enough that temperature keeps low at the handle Self Heating The measurement current of the ins...

Page 6: ...g ON alarm monitoring via display interface and sound Buzzer NO SO alarm monitoring via display interface no sound Buzzer AL LO Alarm setting Min factory setting 5 0 C 200 0 AL K Min alarm limit highest possible value Max Alarm limit AL K Alarm setting Max factory setting 5 0 C AL LO 400 0 Max alarm limit lowest possible value Min Alarm limit ADR Base Address factory setting 01 0 1 9 1 Base addres...

Page 7: ...estarts segment test NOTE If there is no key pressed within the menu mode within 2 minutes the configuration will be cancelled the entered settings are lost 9 Error and System Messages No temperature probe connected Or at GMH 285 BNC cable sensor broken Or at GMH 285 cable sensor shorted ERR 1 Value exceeding measuring range value too high or cable sensor broken ERR 2 Value exceeding measuring ran...

Page 8: ...after switching on are memorized Alarm Freely configurable min max alarm via buzzer display and interface Adjustment Menue offset scale factory settings are recoverable Working temperature 25 to 50 C Storage temperature 25 to 70 C Power supply 9V battery type IEC 6F22 included in scope of supply external d c connector Power consumption 0 20 mA battery life time more than 1500 hours for alkaline ba...

Reviews: