5 | Operation
G 1110
16 / 28
B-H88.0.01.DB2-1.0
3.
The available configuration options are listed for each parameter in the following
representation.
Parameter
Values
Meaning
Zero point correction
PR
.
OF
0
.
0
No offset
PR
.
OF
-5
.
0
..
5
.
0
Offset active
Gradient compensation specification
PR
.
SL
0
No slope
PR
.
SL
-
5
.
00
..
5
.
00
Slope active
Sensor zero point correction offset
PR
.
OF
A zero point shift can be made for the measurement.
Displayed value = measured value
– offset
Standard setting: 0.0, i.e. no correction is made.
Sensor gradient correction
PR
.
SL
The gradient of the measurement can be influenced with this factor (factor is in %):
Displayed value = measured value * (1+Pr.SL/100)
Standard setting: 0.000, i.e. no correction is made.
The changed value is saved and the
adjustment menu
is closed.
STOR
appears in the
display.
NOTE
If the product is switched off without saving the configuration, the last saved values
are reproduced on the next start-up of the product.
Representation
Formula
Outcome of an action