32
11. Rotunda (optional equipment)
❖
This option enables the seat of the treatment table to be rotated up to 330
0
. You
can freeze the set position for treatment using the hand lever which is located
below the seat (s. fig. below). Loosen the hand lever by one rotation of the hand
in clockwise direction, rotate the seat to the desired position and tighten the lever
by turning it in anti-clockwise direction.
WARNING
! Only hold the top of the seat when rotating. Do not grasp
below the metal rotating-disk. RISK OF CRUSHING!
NOTICE!
You cannot use the Rotunda - function when the leg section is
folded by 90
0
. If the leg section contacts the base while rotating, please
fold up the leg section slightly.
12. Initialisation of the software
❖
If a previously-saved position is no longer properly set, you may have to initialize
the software of the SPL med. To correct the differences of position, you have to
carry out the following steps:
I
NITIALIZING
:
1.
The
backrest actuator
must be set to
max
. position. The backrest must have an
angle of approx. 70
0
to the seat! Use
button B
of the relevant manual control
panel / s. fig. below.
2.
The
other actuators
must be set to
min
. position; that means:
a.
The actuator of the work height must be set to its lowest setting! Use
button A1
of the relevant control panel / s. fig. below.