41
NOTICE!
The IRC has no diodes!
You can tell whether the locking function is activated by the diode above
the corresponding button on the manual control panel.
EXTEND
/
RETRACT CASTORS ELECTRICALLY
:
To extend the castors electrically, press the
KEY
button and keep it
pressed until you hear a constant beep.
Now the castors have reached their end position and you can move
the bed.
To retract the castors, proceed in the same way.
Make sure to keep the
KEY
button pressed until you hear a constant
beep indicating that the castors are fully retracted.
WARNING!
The MO1 must always stand on its solid feet. If it stands
on its movable castors, there is RISK OF ROLLING caused by e.g.
leaning against the MO 1. Do not extend the castors if any person is on
the MO 1. Do not move the MO 1 on its castors if any person or load is
on the MO 1; otherwise you damage the substructure or castors.
APPROACHING THE BASIC FUNCTION
:
Press and hold the
SHIFT
button for approximately
3 seconds.
Then press the
START
button and keep it pressed
until the basic
position is reached.