38
➢
Press the buttons M3 and M4 (s. fig. chapter 10.1 M
ANUAL CONTROL PANEL
)
simultaneously. Both diodes will flash and you hear an acoustic signal.
➢
The castors are extracted (final position is reached after approx. 10 seconds and
then a constant beep sounds).
➢
You can move the MO 1 now.
➢
The diodes M3 and M4 flash.
WARNING!
1. RISK OF CRUSHING your toes / feet while lowering or moving the
MO 1.
2.
Do not move the MO 1 now
!
Before moving the MO 1
, unplug it, roll up the power cord properly and store
the power cord in such a way, that you cannot roll over it when moving the MO
1. Do not retract the castors unless the MO 1 stands at its desired site and you
have checked that the cables are not located under the MO 1.
➢
Press and hold the buttons M3 and M4 simultaneously to retract the castors
until you hear a constant acoustic signal. Diodes M3 and M4 go off.
WARNING!
The MO1 must always stand on its solid feet. If it stands
on its movable castors, there is RISK OF ROLLING caused by e.g.
leaning against the MO 1. Do not extract the castors if any person is on
the MO 1. Do not move the MO 1 on its castors if any person or load is
on the MO 1; otherwise you damage the substructure or castors.
NOTICE!
If you press the buttons M3 and M4 although the actuator for
electrically extending and retracting the castors is not present, an error
occurs in the system.
This error manifests itself in a flashing of the diodes above the buttons
and a single, longer beep. The diodes above the buttons continue to
flash and it is then not possible to operate the other buttons of the
manual control panel.
If this error occurs on your MO 1, please contact our customer service.
Summary of Contents for MO 1
Page 14: ...14...
Page 24: ...24 Fig A Fig B Fig C...
Page 46: ...46 Fig shows valve with closed system...
Page 52: ...52 BUTTON ON BUTTON OFF TEMPERATURE SELECTION BUTTON TIME SELECTION BUTTON LEDS A 2 blue LEDs...
Page 89: ...89...