37
10.5.1.1 Overview of temperatures per power level of the heating
Power Level
Approximate temperature
after 1 hour
1
approx. 31 °C
2
approx. 41 °C
3
approx. 50 °C
4
approx. 61 °C
5
approx. 70 °C
6
approx. 74 °C
7
approx. 84 °C
8
approx. 91,5 °C
9
approx. 98 °C
10
approx. 109 °C
10.5.2 Safety switch (interruption of heat generation)
❖
On one side of the
small
heat dome (
5
) is a
SAFETY SWITCH
(
11
/ s. fig. below).
The safety switch prevents heat accumulation between the two heat domes by
interrupting heat generation
if they
are pushed together
.
NOTICE!
Check the safety switch regularly for proper function.
❖
If the two heat domes are pushed together
during a heat application
,
this switch is activated by the pressure exerted by the large heat dome:
heat generation is interrupted
automatically
. The display of the control panel
indicates this interruption
: “
Dome open
“. Above this, the remaining time is
displayed. If you pull out the large heat dome again, the safety switch is
deactivated and the heat generation is continued with the set parameters
(heat level, remaining time, etc.).
5
S
MALL HEAT DOME
(legs / feet)
11
S
AFETY SWITCH