34
10.4 Reset
MLX iDome / MLX i
3
(Cube) Dome
Electric only!
❖
Push the large heat dome towards the foot end until the contact switch is
pressed.
❖
Press and
hold
the arrow buttons
A and B
which are integrated into the frame
on the side of the upper construction, simultaneously for approx. 10 seconds to
reset the software of the MLX iDome / MLX i
3
(Cube) Dome Electric
to its
initial state: recognisable by periodically fast signals (beeps).
❖
After 10 seconds, the signal changes to periodically slower beeps.
❖
Now the two lifting columns must be brought to the lower end position,
alternately with the same arrow buttons
A and B
.
❖
The beep stops and the bed can be raised.
❖
The reset procedure is finished.
Press the arrow buttons
A and B
(working height)
simultaneously for 10 seconds.