31
14.1 Cleaning and care of the surface of the upholstery
❖
The upholstery of the MLK Physio is equipped with a surface made of
ARTIFICIAL LEATHER
.
Clean and care the artificial leather using a special detergent for artificial leather
only
(
optional available at Gharieni; s. chapter 16. O
PTIONALLY AVAILABLE
ACCESSORIES
).
14.2 Cleaning of the housing and of the fittings
Clean the housing and the fittings using a soft damp cloth and a small amount of
soap.
WARNING!
Ensure that water or other liquids cannot get into the
housing and the fittings of the MLK Physio or into the MCPs. RISK OF
ELECTRIC SHOCK! RISK OF FIRE!
15. Malfunctions (possible causes and troubleshooting)
M
ALFUNCTION
P
OSSIBLE CAUSE
(
S
)
T
ROUBLESHOOTING
A
LL ACTUATORS DO NOT WORK
.
1.
No connection to the mains.
1.
Check the connection status.
2.
The power supply cable is damaged.
2.
Power supply cable must be replaced
(please contact the customer
service).
3.
MCP / actuator-plug are not
plugged in properly.
3.
Check proper connection of MCP /
plug.
4.
MCP is defective.
4.
MCP must be replaced (please
contact the customer service).
5.
Fuse in the transformer is defective.
5.
Fuse must be replaced (please
contact the customer service).
6.
Control-box total loss.
6.
Control-box must be replaced
(please contact the customer
service).