11. Malfunctions (causes and troubleshooting)
M
ALFUNCTION
P
OSSIBLE CAUSE
(
S
)
T
ROUBLESHOOTING
A
LL ACTUATORS DO NOT WORK
.
1.
No connection to the mains.
1.
Check the connection status.
2.
The power supply cable is damaged.
2.
Power supply cable must be replaced
(please contact the customer service).
3.
Fuse in the transformer is defective.
3.
Fuse must be replaced (please contact the
customer service).
4.
Control box total loss.
4.
Control box must be replaced (please
contact the customer service).
S
INGLE ACTUATOR DOES NOT WORK
.
1.
Lifting column plug is not connected
properly.
1.
Connect plug properly.
2.
Control box is defective.
2.
Control box must be replaced
(please contact the customer service).
3.
Lifting column cable / plug is defective.
3.
Cable / plug must be replaced
(please contact the customer service).
4.
Lifting column is defective.
4.
Lifting column must be replaced
(please contact the customer service).
NOTICE!
If the MLX is disconnected from the mains while switched on
by pulling the plug or by a power failure, the glass operating panel will
remain black during subsequent start-up.
Disconnect the MLX from the mains and leave the bed disconnected
for 5 minutes. Then plug the mains plug back in.
The MLX will now start up normally and return to normal operation.
If the concept bed shows a malfunction not noted on the schedule above or if
you have checked all items on the above schedule but the problem still exists,
please contact the customer service (s. chapter 11.1 R