GF 263X Series Chlorine Electrodes 

Sensor Preparation

• 

Chlorine sensors are shipped without internal electrolyte solution.

• 

Prior to installation and supplying power, Chlorine sensors must be filled with the appropriate internal electrolyte solution.

• 

Verify the correct electrolyte solution is utilized with the corresponding sensor.

• 

Free Chlorine and Chlorine Dioxide sensors require different electrolyte solutions.

Avoid skin or eye contact with electrolyte solution.

Wear rubber gloves and goggles.

*Safety Data Sheets (SDS) are available online at www.gfps.com.

CAUTION!

 

DO NOT touch the gold tip or the membrane of the sensor.

Initial Fill Procedure:

When adding electrolyte, be prepared for an accidental spill.

Working near a sink is recommended.

1.

  Remove the protective bottle from the end of the electrode.

2.

  Remove the membrane cap from the front of the sensor.

Note:

 When new sensors are shipped, the membrane cap is not tightened to 

the sensor.

3

.  Fill supplied syringe with electrolyte solution.

4.

  Place the electrode on a level surface.

5.

  Insert syringe needle fully into one of the eight electrode holes while injecting 

with electrolyte solution. Slowly injecting the electrolyte solution into the sensor 

to avoid introducing air bubbles. The electrode holds approximately 14 

milliliters of solution. Slowly fill until solution begins to flow out of holes. Do not 

allow the solution to run down the electrode and wet the electrical contacts in 

the DryLoc connector.

6

.  Slowly screw on the membrane cap finger tight. Do not use tools.

 

To avoid damage and contamination, do not touch the white membrane 

surface on the membrane cap.

1

2

3

4

5

6

7

8

9

10

ml

Summary of Contents for GF 3-2630-X

Page 1: ...ly and a secure connection Gold plated contacts and an O ring seal ensure a waterproof and reliable interconnect to the 2650 Amperometric Electronics Integrated temperature element for automatic tempe...

Page 2: ...ts user to risk of potential damage to product by ESD Personal Protective Equipment PPE Always utilize the most appropriate PPE during installation and service of GF products Pressurized System Warnin...

Page 3: ...ignet com for details Declaration of Conformity according to FCC Part 15 This device complies with Part 15 of the FCC rules Operation is subject to the following two conditions 1 This device may not c...

Page 4: ...sink is recommended 1 Remove the protective bottle from the end of the electrode 2 Remove the membrane cap from the front of the sensor Note When new sensors are shipped the membrane cap is not tight...

Page 5: ...le operating pressure must be less than 0 48 bar 7 psi Higher pressures will damage the electrode The electrode should not be used in water containing surfactants oils organic chlorine or stabilizers...

Page 6: ...injecting with electrolyte solution Slowly injecting the electrolyte solution into the sensor to avoid introducing air bubbles The electrode holds approximately 14 milliliters of solution Slowly fill...

Page 7: ...bottle with tap water added Storage periods more than 1 week Remove the membrane cap and internal electrolyte solution Rinse the sensor internal chamber with DI water or cold tap water drain and allow...

Page 8: ...7 11 Fill beaker with a 12 mm inch of the appropriate solution 12 Position or suspend the sensor 6 mm to 12 mm in to in above the appropriate solution DO NOT SUBMERGE THE SENSOR See Figure 2 13 Apply...

Page 9: ...t until fully seated Do Not Use Lubricant or Sealing Tape on Threads Do Not Overtighten Do Not Use Tools pipe tee pipe tee pipe tee pipe tee When using the chlorine panel flow cell it is not necessary...

Page 10: ...tions which help determine the specific application The predominant categories used in disinfection are Free Chlorine Total Chlorine and Chlorine Dioxide Free Chlorine is the sum of chlorine gas Cl2 h...

Page 11: ...resent in water See technical data Surfactants in water Remove surfactants and replace cap Membrane cap coated Clean or replace membrane cap Membrane cap loose Tighten or replace membrane cap pH outsi...

Page 12: ...1 to 20 ppm mg L 3 2632 1 159 001 767 Chlorine Dioxide electrode 0 02 to 2 ppm mg L Accessories and Replacement Parts Mfr Part No Code Description 3 2630 391 159 001 674 Free Chlorine electrolyte 30 m...

Reviews: