Signet 3719 pH/ORP Wet-Tap

Installation in Pipe Sizes 2.5 to 12 Inches

• 

For reliable in-line measurements of pH and ORP, it is 
imperative to position the electrode tip into the process 
stream.

• 

Because of its compact “short stroke” design, the 3719 
requires low-profi le fi ttings to ensure
proper positioning in pipe sizes 
DN65 to DN300 (2.5 to 12 inches).

• 

It is strongly recommended 
to use low-profi le fi ttings 
(customer-supplied).

• 

Choose the 3719 version (-1X or 
-2X) appropriate to the size of the 
branch connection of the required 
fi tting:

 

Use -1X for sizes DN65 to DN100 
(2.5 to 4 in), and -2X for sizes 
DN150 to DN300 (6 to 12 in.)

Installation in Pipe Sizes less than 2.5 Inches

• 

It is possible to install the 3719 into pipe sizes below 
2.5 inches by creating a “fl ow cell” with standard piping 
components.

• 

One simple solution, using a tee fi tting and reducer bushings, 
is shown in the example below.

• 

Many similar confi gurations are conceivable in a wide 
variety of materials, but be very careful to verify dimensional 
compatibility.

• 

Select an appropriate installation orientation to avoid the 
entrapment of air inside the fl ow cell.

• 

Contact your local Georg Fischer Sales and Support offi  ce 
for assistance.

3-3719-1X
pH/ORP Wet-Tap
(Fully Inserted)

2 in /d50 mm
Tee
(S x S x S)

1½ in. 
Threads

Flush Style Reducer
Bushings (Spg x S)
 

2 in./d50 x ½ in./12 mm

 

2 in./d50 x ¾  in./20 mm

 

2 in./d50 x 1 in./25 mm 

 

2 in./d50 x 1¼ in./30 mm

 

2 in./d50 x 1½ in./40 mm

3-3719-1X
pH/ORP Wet-Tap

Flush Style Reducer Bushing 
(Spg x FT)  2 in. x 1½ in. 
 

250 mm 
(9.9 in.) 

45 mm 
(1.75 in.) 

115 mm
(4.5 in.)

218 mm 
(8.57 in.) 

OR

89.4 mm

(3.5 in.)

• 

Provide 20 inches (minimum) clearance from the top of the 
pipe for electrode removal..

Installation continued

Installation continued

Summary of Contents for Signet 3719

Page 1: ...omers to read the speci cations carefully before installing and operating this product Improper use can cause components and process liquids to be expelled at high speeds and cause serious personal in...

Page 2: ...s or application failures e g high temperature chemical poisoning dry out or mishandling e g broken glass damaged membrane freezing and or extreme temperatures Thank you for purchasing the Signet line...

Page 3: ...ee ordering information on page 8 DryLoc refers to the electrode connector style d 3 2751 1 or 2760 DryLoc pH ORP Sensor Electronics Preampli er e 3 2751 2 electronics NOTE Output signal options See s...

Page 4: ...g a ow cell with standard piping components One simple solution using a tee tting and reducer bushings is shown in the example below Many similar con gurations are conceivable in a wide variety of mat...

Page 5: ...peration Use an appropriate thread sealant to prevent leaks The piping system can now be safely pressurized Inspect the installation for leaks 3 4 Electrode Piston Electrode Installation Before instal...

Page 6: ...ean any scaling and debris found on or in the piston tip and surrounding area 2 3 4 5 1 5 3 4 Turn the locking shroud turn counterclockwise and lift up to remove it from the Wet Tap assembly 5 Remove...

Page 7: ...the tip of the piston and into each O ring groove 6 Lubricant 7 countersink hole Stainless Steel set screw hole Locate the countersink hole on the nosepiece Verify that it is clean and free of debris...

Page 8: ...PT 3 3719 12 159 000 806 pH ORP Wet Tap ISO 7 1 R 1 5 3 3719 22 159 000 807 Wet Tap Assembly ISO 7 1 R 2 Parts and Accessories 3 2756 WTP 159 001 390 Electrode pH DryLoc plastic bulb Pt1000 Wet Tap 3...

Reviews: