GF Signet 2850 Operating Instructions Manual Download Page 5

5

2850 Conductivity/Resistivity Sensor Electronics

4 to 20 mA Wiring

Digital (S

3

L) Wiring

Dual Digital (S

3

L) Input Wiring

Dual Input is available on Digital (S

3

L) output 

model 3-2850-63 only, for use with the Signet 
8900 Multi-Parameter Controller.

Sensors may be the same or different cell 
constants.

If the Dual Input model is used with a single 
sensor, connect the sensor to Input 1 and set 
all of the Input 2 switches to OFF (OPEN).

The Digital (S

3

L) output is compatible with the 

Signet 8900 Multi-Parameter Controller and 
the Signet 9900 Transmitter.

Refer to the wiring sections of the 8900 or 9900 
manual to determine cable length limitations.

IMPORTANT:  Provide Earth Ground as 

indicated by 

 symbol.

Maximum length of 4 to 20 mA loop is 300 meters (1000 ft)

SW1 

D3 

ON 

CTS 

ON 

CTS 

4 20 

V- 

V+ 

Temp. IN 
(WHITE) 

Sensr Gnd 
(SHIELD) 

Signal IN 
(RED) 

Signal RTN 
(BLK) 

9 VDC 

to

26 VDC

4-20 mA Loop Output

SW1 

D3 

1

2

3

ON 

CTS 

3-8900.621C 

I/O Module 3-8900.401-X 

10 

11 

12 

13 

14 

+5VDC (Black) 

Freq. Input (Red) 

GND (Shield) 

+5VDC (Black) 

Freq. Input 2 (Red) 

S  L (Red) 

GND (White/Shield) 

+5VDC (Black) 

S  L (Red) 

GND (White/Shield) 

 

Analog Output 1 

 

Analog Output 2 

(if applicable) 

(if applicable) 

Frequency 

Input 

Frequency 

Input 2 

OR 

S3L 

Input 

 

S3L 

Input 

V+

DA
T

A

GND

SHLD

V+

DC Power

Loop V

oltage

3-9900.395

H COMM Module

Gnd

Data

Temp. IN
(WHITE)

Sensr Gnd
(SHIELD)

Signal IN
(RED)

Signal RTN
(BLK)

Black

Red

White

9900 S

3

L Input

SW1 

D3 

ON 

CTS 

ON 

CTS 

V+ 

3-8900.621C 

I/O Module 3-8900.401-X 

10 

11 

12 

13 

14 

+5VDC (Black) 

Freq. Input (Red) 

GND (Shield) 

+5VDC (Black) 

Freq. Input 2 (Red) 

S  L (Red) 

GND (White/Shield) 

+5VDC (Black) 

S  L (Red) 

GND (White/Shield) 

 

Analog Output 1 

 

Analog Output 2 

(if applicable) 

(if applicable) 

Frequency 

Input 

Frequency 

Input 2 

OR 

S3L 

Input 

 

S3L 

Input 

Gnd

Data

Input 1

Input 2

Sensor 1

Sensor 2

Note:  

The 2819-2823 and 2839-2842 sensors can 
connect directly to the 9900 via the 9900's 
optional Direct Conductivity/Resistivity 
Module 3-9900.394 (159 001 699).

Dual Input applications usually include one sensor with a smaller cell constant and 
one sensor with a larger cell constant. 

Place the larger cell constant on Input 1.

This allows periodic recalibration of the larger cell by simply turning Input 2 OFF (OPEN).

IMPORTANT

:  Provide Earth Ground as indicated by 

 symbol.

Summary of Contents for Signet 2850

Page 1: ...vity test solution values for simple field calibration The Conductivity Sensor provided with integral systems will have its custom cell constant information programmed into the electronics at the fact...

Page 2: ...t accept any interference received including interference that may cause undesired operation Specifications Caution Warning Danger Indicates a potential hazard Failure to follow all warnings may lead...

Page 3: ...li Cleaners Acids Bases 0 01 Cell 0 10 Cell 1 0 Cell 10 0 Cell 20 0 Cell 3 2841 3 2842 26 8 mm 1 13 in 41 9 mm 1 65 in 3 2823 203 mm 8 0 in 89 mm 3 5 in 3 2839 59 3 mm 2 33 in 73 7 mm 2 90 in 3 2840 2...

Page 4: ...sanitary flange fittings must be installed using this option While the 2850 5X and 2850 6X electronics cannot be submerged either model will accommodate tank installation Select any electrode with a...

Page 5: ...q Input 2 Red S L Red GND White Shield 5VDC Black S L Red GND White Shield 3 3 Analog Output 1 Analog Output 2 if applicable if applicable Frequency Input 1 Frequency Input 2 OR S3L Input 2 S3L Input...

Page 6: ...W3 or SW4 Integral 2850 Systems Single Input The 2850 51 XX and 2850 52 XX are shipped from the factory with the custom cell constant and temperature offset programmed into the single input electronic...

Page 7: ...to 500 S 0 to 1000 S O C C C 10 to 2 M 5 to 0 S 50 to 0 S 500 to 0 S 1 000 to 0 S C O C O 0 to 2 M 0 to 10 S 0 to 100 S 0 to 1 000 S 0 to 2 000 S C O C C 2 to 0 M 10 to 0 S 100 to 0 S 1 000 to 0 S 2...

Page 8: ...ed if both inputs are used or if Input 2 is disabled To disable Input 2 Set SW4 4 to OPEN Place each sensor into a test solution that represents the operating range of that sensor When SW1 is pressed...

Page 9: ...e test solution is outside the 10 tolerance The electronics or the electrode is defective Use fresh test solutions and restart the calibration Replace the electronics or the electrode After completing...

Page 10: ...1 Turn the power off 2 Set SW3 to accept a 0 01 cell constant Leave SW4 in the all OPEN position 3 Install the simulator into the 2850 at the Input 1 conductivity electrode terminals 4 Connect the 285...

Page 11: ...11 2850 Conductivity Resistivity Sensor Electronics Notes...

Page 12: ...tem Digital S3 L output 0 01 cell ISO threads 3 2850 51 40D 159 001 344 Integral 2850 system Digital S3 L output 0 1 cell ISO threads 3 2850 51 41D 159 001 345 Integral 2850 system Digital S3 L output...

Reviews: