GF Signet 2850 Operating Instructions Manual Download Page 2

2

2850 Conductivity/Resistivity Sensor Electronics

General

Compatible Electrodes ............. Any Signet Conductivity/

Resisitivity electrode 

(2818-2823, 2839-2842)

Temperature Compensation .....PT-1000 RTD (2% per °C)

Pure Water Compensation ....... Auto-switching when using 

0.01 cell and raw conductivity 

value < 0.5 

μ

S

EasyCal

Automatic recognition of the following conductivity values:

 •  146.93 

μ

S, 1408.8 

μ

S, 12856 

μ

S (@ 25 °C)

  (Test solutions per ASTM D1125-95)

 •  10 

μ

S, 100 

μ

S, 200 

μ

S, 500 

μ

S, 1000 

μ

S, 5000 

μ

S, 10,000 

μ

S, 

50,000 

μ

S, 100,000 

μ

S (@ 25 °C)

Power

 •  12 to 24 VDC (4 to 20 mA output)

 •  5 VDC ±5% regulated, 3.0 mA max. for Digital (S

3

L) output 

(Reverse polarity and short circuit protected)

Digital (S

3

L) Output

 ................. Serial ASCII, TTL level 9600 bps

Accuracy

 • Conductivity ..........................±2% of reading

 • Temperature ..........................±0.5 °C
Resolution

 • Conductivity ..........................0.1% of reading

 • Temperature ..........................< 0.2 °C
Update Rate

 • Single Input models ..............< 600 ms

 • Dual Input models .................< 1.2 s

Current Output

Field-selectable ranges. Factory set span:

 • 0.01 cell (2818, 2819, 2839) ..4 to 20 mA = 0 to 100 

μ

S

 • 0.10 cell (2820, 2840) ...........4 to 20 mA = 0 to 1000 

μ

S

 • 1.0 cell (2821, 2841) .............4 to 20 mA = 0 to 10,000 

μ

S

 • 10.0 cell (2822, 2842) ...........4 to 20 mA = 0 to 200,000 

μ

S

 • 20.0 cell (2823) .....................4 to 20 mA = 0 to 400,000 

μ

S

Max. Loop Resistance ..............   50 

Ω

 @ 12 VDC

325 

Ω

 @ 18 VDC

600 

Ω

 @ 24 VDC

Accuracy ...................................±2% of output span
Resolution.................................7 

μ

A

Update Rate .............................<600 ms
Error Indication .........................22 mA

Available data via Digital (S

3

L) Output

 • Raw conductivity

 • Calibrated conductivity

 • Calibrated temperature-compensated conductivity

 • Temp. Error Indication ........... Open input and out-of-range 

diagnostics for temperature or 

internal electronic error.

Environmental

Material .....................................PBT

Enclosure Rating ......................NEMA 4X/IP65

Operating Temperature .............-10 to 85 °C (14 to 185 °F)

Storage Temperature ................-20 to 85 °C (-4 to 185 °F)

Relative Humidity......................0 to 95%, non-condensing

Shipping Weight .......................0.75 kg (1.75 lb.)

Standards and Approvals

   •  CE, RoHS Compliant

   •   Manufactured under ISO 9001 for Quality, 

ISO 14001 for Environmental Management and 

OHSAS 18001 for Occupational Health and Safety.
China RoHS (Go to www.gfsignet.com for details)
 This device complies with Part 15 of the FCC rules. 

Operation is subject to the following two conditions:

(1) This device may not cause harmful interference, and 

(2)  This device must accept any interference received, 

including interference that may cause undesired operation.

Speci

fi

 cations

Caution / Warning / Danger

Indicates a potential hazard. Failure to follow all warnings 
may lead to equipment damage, injury, or death

Electrostatic Discharge (ESD) / Electrocution Danger

Alerts user to risk of potential damage to product by ESD, 
and/or risk of potential of injury or death via electrocution.

Personal Protective Equipment (PPE)

Always utilize the most appropriate PPE during 
installation and service of Signet products.

Pressurized System Warning

Sensor may be under pressure, take caution to vent 
system prior to installation or removal. Failure to do so 
may result in equipment damage and/or serious injury.

Note / Technical Notes

Highlights additional information or detailed procedure.

Refer to your local Georg Fischer Sales of

fi

 ce for the most 

current warranty statement.
All warranty and non-warranty repairs being returned must 
include a fully completed Service Form and goods must be 
returned to your local GF Sales of

fi

 ce or distributor. 

Product returned without a Service Form may not be 
warranty replaced or repaired.
Signet products with limited shelf-life (e.g. pH, ORP, chlorine 
electrodes, calibration solutions; e.g. pH buffers, turbidity 
standards or other solutions) are warranted out of box but not 
warranted against any damage, due to process or application 
failures (e.g. high temperature, chemical poisoning, dry-out) or 
mishandling (e.g. broken glass, damaged membrane, freezing 
and/or extreme temperatures).

Thank you for purchasing the Signet line of Georg Fischer 
measurement products.
If you would like to register your product(s), you can now 
register online in one of the following ways:
 •  

Visit our website 

www.gfsignet.com. 

Under 

Service and Support

 click on 

Product Registration Form

  •  If this is a pdf manual (digital copy), 

click here

Warranty Information

Product Registration

Safety Information

  1.  Depressurize and vent system prior to installation or removal.
  2.  Con

fi

 rm chemical compatibility before use.

  3.  Do not exceed maximum temperature/pressure speci

fi

 cations.

  4.  Wear safety goggles or faceshield during installation/service.
  5.  Do not alter product construction.
  6.  When using chemicals or solvents, care should be taken 

and appropriate eye, face, hand, body, and/or respiratory 
protection should be used.

Summary of Contents for Signet 2850

Page 1: ...vity test solution values for simple field calibration The Conductivity Sensor provided with integral systems will have its custom cell constant information programmed into the electronics at the fact...

Page 2: ...t accept any interference received including interference that may cause undesired operation Specifications Caution Warning Danger Indicates a potential hazard Failure to follow all warnings may lead...

Page 3: ...li Cleaners Acids Bases 0 01 Cell 0 10 Cell 1 0 Cell 10 0 Cell 20 0 Cell 3 2841 3 2842 26 8 mm 1 13 in 41 9 mm 1 65 in 3 2823 203 mm 8 0 in 89 mm 3 5 in 3 2839 59 3 mm 2 33 in 73 7 mm 2 90 in 3 2840 2...

Page 4: ...sanitary flange fittings must be installed using this option While the 2850 5X and 2850 6X electronics cannot be submerged either model will accommodate tank installation Select any electrode with a...

Page 5: ...q Input 2 Red S L Red GND White Shield 5VDC Black S L Red GND White Shield 3 3 Analog Output 1 Analog Output 2 if applicable if applicable Frequency Input 1 Frequency Input 2 OR S3L Input 2 S3L Input...

Page 6: ...W3 or SW4 Integral 2850 Systems Single Input The 2850 51 XX and 2850 52 XX are shipped from the factory with the custom cell constant and temperature offset programmed into the single input electronic...

Page 7: ...to 500 S 0 to 1000 S O C C C 10 to 2 M 5 to 0 S 50 to 0 S 500 to 0 S 1 000 to 0 S C O C O 0 to 2 M 0 to 10 S 0 to 100 S 0 to 1 000 S 0 to 2 000 S C O C C 2 to 0 M 10 to 0 S 100 to 0 S 1 000 to 0 S 2...

Page 8: ...ed if both inputs are used or if Input 2 is disabled To disable Input 2 Set SW4 4 to OPEN Place each sensor into a test solution that represents the operating range of that sensor When SW1 is pressed...

Page 9: ...e test solution is outside the 10 tolerance The electronics or the electrode is defective Use fresh test solutions and restart the calibration Replace the electronics or the electrode After completing...

Page 10: ...1 Turn the power off 2 Set SW3 to accept a 0 01 cell constant Leave SW4 in the all OPEN position 3 Install the simulator into the 2850 at the Input 1 conductivity electrode terminals 4 Connect the 285...

Page 11: ...11 2850 Conductivity Resistivity Sensor Electronics Notes...

Page 12: ...tem Digital S3 L output 0 01 cell ISO threads 3 2850 51 40D 159 001 344 Integral 2850 system Digital S3 L output 0 1 cell ISO threads 3 2850 51 41D 159 001 345 Integral 2850 system Digital S3 L output...