3.3 Connections
Connections
Cold Water
Water quality
Water quality Drinking water with max
5°dH without softener
Water with 5°dH-35°dH with softener.
Connection
G 1/2”
Pressure
100-800 kPa
Flow rate
min 20 l/min
Hot Water
Water quality
Preheated drinking water with a max
5°dH 45°C-60°C (113°F-140°F) without
softener.
Water with 5°dH-35°dH 45°C-60°C
(113°F-140°F) with softener.
Connection
G 1/2”
Pressure
100-800 kPa
Flow rate
min 20 l/min
Treated water
Temperature
max 60°C (140°F)
Connection
G 1/2”
Pressure
100-800 kPa
Flow rate
min 20 l/min
The maximum conductivity is
determined by the desired cleaning
results.
Dosage
Pump accuracy
±10%
Drainage
Connection x2
Ø 20 mm inside
Ø 25 mm outside
Capacity
50 l/min
Residual volume in the
chamber
Max. 0.5 liters.
Electrical connections
See section
.
3 TECHNICAL DATA