31 CS100/CS300 BATTERY OPERATION
CS100/100i Preventive Maintenance
Two preventive maintenance schedules have been provided.
Schedule A indicates which actions should be taken by either the Clinical User or by a Biomedical Technician (BMET). These
steps do not require the use of tools and may be performed in a clinical setting.
Schedule B indicates the actions which should be performed only by a BMET or other qualified service personnel. Tools are
required and in some cases the instrument covers must be removed.
WARNING: Preventive Maintenance should not be performed when the IABP is attached to a patient.
Schedule B
To be performed by the BMET.
INTERVAL
Required Action
(REFER TO SERVICE MANUAL)
Every 6
Months
Every
2500
Hours
Every
5000
Hours
Check battery for rated voltage and
check battery run time. Replace
batteries when operating time is
marginal or after three (3) years.
•
* This does not imply a three-year warranty.
INTERVAL
Required Action
Before or After
Each Use
Every
Month
Every 6
Months
Every
1000 Hrs.
of Use or
2 Years
Check battery system
Check Battery Indicator
•
Check battery run time. Replace
batteries when operating time is
marginal.
•
* Whichever comes first.
Schedule A
To be performed by the clinical user or the BMET.
CS100 Battery
Preventive Maintenance