21 CS100/CS300 BATTERY OPERATION
CS300 Preventive Maintenance
Two preventive maintenance schedules have been provided.
Schedule A indicates which actions should be taken by either the Clinical User or by a Biomedical Technician IBMET). These steps do not
require the use of tools and may be performed in a clinical setting.
Schedule B indicates the actions which should be performed only by a BMET or other qualified service personnel. Tools are required and in
some cases the instrument covers must be removed.
WARNING: Preventive Maintenance should not be performed when the IABP is attached to a patient.
CAUTION: This product requires scheduled preventative maintenance in order to maintain its specified performance. Note that
maintenance includes periodic cleaning to assure that proper cooling airflow of the system‘s electronics is maintained.
Schedule B
To be performed by the BMET.
INTERVAL
Required Action
(REFER TO SERVICE MANUAL)
Every 6
Months
Every
2500
Hours
Every
5000
Hours
check battery for rated voltage and
check battery run time. Replace
batteries when operating time is
marginal or after three (3) years.*
•
* This does not imply a three-year warranty.
INTERVAL
Required Action
Before or After
Each Use
Every
Month
Every 6
Months
Every
1000 Hrs.
of Use or
2 Years*
Before each use, ensure that the
Battery Charging LED is illuminated;
or, if batteries are in use, that the
Battery Icon shows a full battery. See
page 7 of this document.
•
* Whichever comes first
Schedule A
To be performed by the clinical user or the BMET.
CS300 Battery
Preventive Maintenance