CS100/CS300 Battery Operation

 

Quick Reference Guide 

Summary of Contents for CS100

Page 1: ...ated guide for proper battery operation and maintenance for the CS100 and CS300 For complete product instructions please refer to the CS100 and CS300 Operating Instructions CS100 CS300 Battery Operation Quick Reference Guide ...

Page 2: ... accompany the CS100 and CS300 Intra aortic Ballon Pumps The CS100 and CS300 batteries are designed to be used during transport or portable operation and it is critical to keep these batteries charged at all times to avoid interruption in power during transport Additionally it is important to maintain the batteries in accordance with CS100 and CS300 Operating Instructions CS100 CS300 ...

Page 3: ...s Table of Contents Section Page Battery Overview 4 6 Battery Status 7 14 CS300 Battery Operation 15 20 CS300 Battery Preventive Maintenance 21 CS100 Battery Operation 22 30 CS100 Battery Preventive Maintenance 31 Key Battery Warnings and Cautions 32 ...

Page 4: ... 25 hours minimum from full charge 120 bpm 22 5 C Approximately 3 0 hours typical with new battery from full charge 90 bpm 22 5 C NOTE A reduction in runtime can occur over a battery s life for reasons such as age storage temperature and discharge depth Recharge Time 18 hours minimum from complete discharge to full charge 22 5 C 8 hours to achieve at least 90 percent charge typically ...

Page 5: ...r electrolyte The batteries are sealed to prevent leakage from the terminals or case Replace batteries as required Batteries should be replaced after 100 full discharge cycles at no more than three 3 year intervals or if run time is less than the minimum run time of 135 minutes minimum at 120 bpm Disposal of batteries should be conducted in accordance with local statutes and the labeling shown on ...

Page 6: ...s a safety vent that will allow gases to be released should its rating be exceeded If this vent is actuated do not operate the unit and contact a qualified service representative Do not short circuit the battery or battery pack terminals Keep battery pack covers on at all times to avoid objects or liquids falling on batteries Excessive heat is very detrimental to battery life Do not operate the sy...

Page 7: ...an active AC Mains source the AC Mains switch is On and the battery charging status LED is illuminated This charging condition must be maintained even when the system is not in use Note that 18 hours of charging is required to fully charge the batteries FOR CS300 CAUTION The system should be configured to charge the internal batteries whenever AC power is available Hence keep the system connected ...

Page 8: ...n 30 minutes approximately of internal battery life remains When this message appears the battery symbol is displayed as empty and starts flashing See the Troubleshooting and User Maintenance Chapters in the Operating Instructions for additional information TRIGGER HEART RATE bpm MEAN mmHg AUG mmHg AUG ALARM mmHg SYSTOLIC mmHg DIASTOLIC mmHg IAB STATUS PRESSURE SOURCE IAB FILL MODE SLOW GAS ALARM ...

Page 9: ...alarm via a specific text message Textual information is legible at the operator s position up to 1 meter from the front of the display panel Additionally alarms are further complimented by an option to display context sensitive help information to aid in the understanding and guide in the resolution of the alarm condition Alarm and advisory messages are grouped into the following categories in or...

Page 10: ...ay indicate a need for corrective action A yellow flashing alarm icon with two 2 exclamation points denotes Medium Priority Alarms All Medium Priority Alarms have a uniform audio tone The combination of three 3 notes for Medium Priority Alarms is played in the following sequence three notes apause and then this cycle repeats NOTE A Low Battery is categorized as a Medium Priority Alarm Medium Prior...

Page 11: ...e help screens for alarm messages are only available while an alarm condition or informational message exists The following is the associated help screen for the battery related Medium Priority Alarm LOW BATTERY Associated Help Screen Probable Cause Corrective Actions There is less than 30 minutes of battery operating time remaining 1 Verify MAINS power switch located above the AC power cord conne...

Page 12: ...Messages BATTERY IN USE Associated Help Screen Probable Cause Corrective Actions The IABP is being powered from the internal batteries 1 Verify that it is necessary to operate from the internal battery If appropriate switch to the AC power source 2 Verify MAINS power switch located above the AC power cord connector is ON CAUSE AND RESPONSE Operation Trigger Mode Trigger Source Detailed Cause Syste...

Page 13: ...ery alarm NOTE If a Low Battery alarm occurs immediately connect the system to AC power CAUSE AND RESPONSE Operation Mode Trigger Source Detailed Cause System Response Reset All All The Battery Test Due Date or Battery Replacement Date predate the current system date at startup or the internal battery has a total accumulated discharge time in excess of 100 total discharge cycles None At startup th...

Page 14: ...ry run time may be reduced Monitor the system for a Low Battery alarm NOTE If a Low Battery alarm occurs immediately connect the system to AC power CAUSE AND RESPONSE Operation Mode Trigger Source Detailed Cause System Response Reset All All The internal battery failed to reach a satisfactory level of charge within 10 hours None At startup the message is cleared when the START key is pressed Durin...

Page 15: ... unit if AC power is not available CS300 Hospital Cart The IABP console is mounted on a hospital cart for intra hospital transport The console may be separated from the cart to accommodate very tight spaces within a hospital room venue All four 4 casters on the hospital cart swivel to facilitate movement in any direction The two casters closest to the cart handle are multifunctional They can be lo...

Page 16: ... filled pressure monitoring catheters Altitude changes are compensated for automatically in the Autofill mode or manually in the Manual Fill mode 3 The system must be properly secured when used in an ambulance helicopter or fixed wing aircraft The Universal Transport System UTS is specifically designed for air and vehicular transport The UTS version mechanically attaches to a docking station for h...

Page 17: ...re not connected will stop the therapy i e power down the pump CAUTION When AC power operation is intended insure that the system is plugged into AC mains that the mains switch is in the On position and that the Battery in Use informational message is NOT displayed CAUTION System batteries must be properly maintained and periodically tested CAUTION The system should be configured to charge the int...

Page 18: ...g the following occurs The Medium Priority Alarm tone is activated The Low Battery Medium Priority Alarm message is displayed on the screen continuously The Battery Indicator is displayed as empty and it starts flashing The condensate removal module will not operate to conserve remaining power A reduction in run time will occur over a battery s life due to age storage temperature and discharge cyc...

Page 19: ...e message Battery in Use will not be displayed during proper AC inverter operation Specifications for DC Source and Inverter for Use with the IABP Voltage Output 100 120 220 240 VAC l 0 Frequency 50 Hz 2 Hz 60 Hz 2 Hz Overshoots Does not continuously generate overshoots greater than 375 Volts peak with widths greater than 10 mSec when powering the system Waveform Sine wave or modified sine wave Ou...

Page 20: ...ng the pump console to the cart Attach monitor to top of pump console Extend console handle until it locks tilt pump and pull to transport Unlock console by pressing tab to right of console release handle and pull handle straight out Lift pump console straight up and off cart Detach monitor from cart handle by pressing button on rear of monitor Release button PULL 2 3 4 1 CAUTION When transporting...

Page 21: ...r to maintain its specified performance Note that maintenance includes periodic cleaning to assure that proper cooling airflow of the system s electronics is maintained Schedule B To be performed by the BMET INTERVAL Required Action REFER TO SERVICE MANUAL Every 6 Months Every 2500 Hours Every 5000 Hours check battery for rated voltage and check battery run time Replace batteries when operating ti...

Page 22: ... direction The two casters closest to the cart handle are multifunctional They can be locked in the straight position directional lock or used as a brake total lock The two casters on the side opposite the handle have total lock only NOTE The casters on the handle side of the cart may or may not have a braking feature Refer to markings on the casters It is recommended that all four casters are loc...

Page 23: ...available in two versions with identical clinical functionality A console version with a hospital cart or the CS100 Universal Transport System UTS is specifically designed for air and vehicular transport The UTS version mechanically attaches to a docking station for high strength mounting WARNING The user should continually rely on visual alarm messages during high noise transport situations The F...

Page 24: ...ttery In Use status message and the Battery Indicator are displayed when the CS100 is operating from the internal rechargeable battery When the battery has approximately 30 minutes of operating time remaining the following occurs An audible double beep alarm is activated for 30 seconds The Low Battery or Low Battery EXT CS l 00i only alert messages are displayed on the screen continuously The Batt...

Page 25: ... l 8 hours to fully charge a low battery Allow 8 hours to achieve al least 90 percent charge typically 4 A fully charged new internal battery will provide at least 135 minutes of portable operation NOTE A reduction in run time will occur over a battery s life due to age storage temperature and discharge depth Batteries which are continually subjected to complete discharge cycles without the recomm...

Page 26: ... instructions 3 Verify that the BATTERY IN USE advisory message and the Battery Indicator are displayed NOTE Battery charging is not active in this state 4 When AC power is restored the system automatically reverts from internal battery operation to AC Mains usage The internal battery pack will resume charging while the system operates from AC Mains power Always verify that the Battery Charging LE...

Page 27: ...ernal battery pack 1 Connect a voltage compatible external source to the External DC Input connector 2 Internal battery will not be charged in this mode 3 The IABP ON OFF switch will activate the system 4 Interruption of the external DC source power will result in portable internal battery operation 5 The internal BATTERY INDICATOR will not be displayed during external DC operation CS100 Battery O...

Page 28: ...ill not be displayed during proper AC inverter operation Specifications for DC Source and Inverter for Use with the CS100 Voltage Output 100 120 220 240 VAC 10 Frequency 50 Hz 2 Hz 60 Hz 2 Hz Overshoots Does not continuously generate overshoots greater than 375 Volts peak with widths greater than 10 mSec when powering the system Waveform Sine wave or modified sine wave Output Capability Minimum of...

Page 29: ...gency Battery Back Up Recommendations Datascope recommends that a back up to the internal battery always be available WARNING Replacing the internal battery when AC mains are not connected will stop the therapy i e power down the pump CS100 Battery Operation ...

Page 30: ...re that the wheels of the cart are in the locked position when removing the pump console from the cart or returning the pump console to the cart Attach monitor to top of pump console Extend console handle until it locks tilt pump and pull to transport Unlock console by pressing tab to right of console release handle and pull handle straight out Lift pump console straight up and off cart Detach mon...

Page 31: ...aintenance should not be performed when the IABP is attached to a patient Schedule B To be performed by the BMET INTERVAL Required Action REFER TO SERVICE MANUAL Every 6 Months Every 2500 Hours Every 5000 Hours Check battery for rated voltage and check battery run time Replace batteries when operating time is marginal or after three 3 years This does not imply a three year warranty INTERVAL Requir...

Page 32: ...he internal batteries whenever AC power is available Hence keep the system connected to a live AC receptacle with the Mains On Off switch in the On position If the system is to be stored for extensive periods of time or in ambient temperatures above the operating range see the Battery Section of the User Maintenance Chapter in the Operating Instructions for additional information CAUTION Prior to ...

Page 33: ......

Page 34: ...tinge AB its subsidiaries or affiliates in the United States or other countries Maquet is registered with the U S Patent and Trademark Office STATLOCK is a registered trademark of C R Bard Inc Copyright 2019 Datascope Corp All rights reserved CAUTION Federal U S A law restricts this device to sale distribution and use by or on the order of a physician Refer to Instructions for Use for current indi...