4.2.4 PATIENT CIRCUIT TEST
The test measures the compliance and
resistance of the patient breathing system.
The patient circuit resistance is automatically
measured to determine if the ventilator
maintains the specified accuracy with the
connected breathing circuit.
1.
Connect the complete breathing system
to be used on the patient. If an active
humidifier is used, it must be filled with
water.
S
VX-155a_XX
Important: When blocking the end of
the complete breathing system, make
sure there is no leakage. Leakage will
affect the circuit compliance
compensation calculation.
2.
Block the end of the complete breathing
system and follow the on-screen
instructions. The circuit compliance and
resistance are automatically measured.
Go to Compensate for Circuit compliance
S
VX-155b_XX
3.
Unblock the end of the complete
breathing system and follow the on-screen
instructions.
SERVO-s VENTILATOR SYSTEM V7.0, User's Manual
45
|
Operation overview
| 4 |
Summary of Contents for Maquet SERVO-s VENTILATOR SYSTEM V7.0
Page 1: ...x User s Manual SERVO s VENTILATOR SYSTEM V7 0 ...
Page 2: ......
Page 4: ...4 SERVO s VENTILATOR SYSTEM V7 0 User s Manual Infologic 1 64 2 1 TABLE OF CONTENTS ...
Page 14: ...14 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 1 Introduction ...
Page 40: ...40 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 3 Power supply ...
Page 66: ...66 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 5 Monitor and record ...
Page 118: ...118 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 6 Ventilation modes and functions ...
Page 162: ...162 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 11 Technical data ...
Page 176: ...176 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 13 Appendix User interface ...
Page 182: ...182 SERVO s VENTILATOR SYSTEM V7 0 User s Manual 15 Index ...
Page 183: ...x ...
Page 185: ...SERVO s VENTILATOR SYSTEM V7 0 User s Manual ...