34
LRG 1x-6x - Installation & Operating Manual -
819878-00
Conductivity measurement and continuous blowdown control
Installation of the conductivity electrode via a measuring vessel with connection of a continuous blow-
down valve.
Key, see page 36
* Minimum distances (R):
■
■
LRG 16-60/LRG 17-60
R = 30 mm
■
■
LRG16-61
R = 60 mm
Fig. 15
Installation examples with dimensions
All lengths and diameters in mm
NW
LW
NB
R*
A
O
K
!
+
-
600
338
~ 250
197
G1
1
2
3
Sealing ring
D 33 x 39
DIN 7603-2.4068,
bright annealed
e.g. LRG 16-60
4
Distance (A), depending on connection flange:
■
■
DN 15 mm
A = 182 mm
■
■
DN 20 mm
A = 184 mm
■
■
DN 25 mm
A = 184 mm
■
■
DN 40 mm
A = 189 mm