Trouble
Possible Causes
Remedy
Low velocity air
Low voltage on house current
Measure voltage – contact
certification company
Filter loaded
Replace filters – contact
certification company
Improper adjustment of blower
motor
Adjust to proper reading on
pressure gauge – contact
certification company
Obstruction in plenum
Remove obstruction
Leak between blower and
plenum
Repair leak – contact certification
company
High velocity air
Filter rupture
Replace Filters – Contact
certification company
Supply filter gasket leak
Find and repair leak – contact
certification company
Exhaust air outlet obstruction
Remove obstruction
Improper adjustment of blower
motor speed control
Adjust to proper reading on
pressure gauge – contact
certification company
Non-laminar airflow
Large object obstructing airflow
Remove item
Large leak in supply filter
Replace Filters – Contact
certification company