User Manual
Spirostik
SPIROSTIK_MAN_ENG_V1.2.2_REV01.DOCX
Page 8
2.2.2
Safe Use of Spirostik
Spirostik is part of a PC based medical device. It is highly recommend
that the user observes all warning notices in order to ensure a safe
use of the device.
The Spirostik complete should not be used in conjunction with
multiple connection plug sockets or extension leads.
Spirostik must only be used in combination with peripheral
devices which fulfill the Norm DIN EN 60950 for office devices or
the norm DIN EN 60601 for medical electrical devices. Always
ensure the minimum requirements for the computer are met
(see technical specification). Non-Medical devices should not be
used close to a patient always ensure a minimum distance of
1.5m between patient and non-medical computer systems.
Flow sensors and pressure tubes must not be cleaned! Humidity
inside the pressure tube may affect the accuracy of the
measurement. Replace the pressure tube in case of
contamination. Replace the flow sensor after each patient.
Always observe instructions in the disinfection and cleaning
section.
The User must not touch the patient and the signal output of the
computer. The device should not be used in association with a
mobile power strip or extension cable.
Large quantities of saliva and/or. Humidity in the flow sensor may
impair the flow measurement. The flow sensor should be
replaced in these conditions. In addition extreme temperature as
well as extreme altitude may impair the flow measurement.
Consider the site conditions indicated in the technical data.
This device must not be modified without the approval of the
manufacturer.
Power supply can be disconnected by releasing the USB plug
connection.