Instructions for Use
Ergostik Blueflow
Serial no.: xx|8|201|yyy and 2203xxxxx
Page 64
Version: 6 | Release date: 02 August 2022
7.5
Calibrating Ergostik Blueflow
There are various processes to be carried out for Ergostik
Blueflow in order to calibrate or validate the volume and gas
measurement.
The required intervals can be found in the tables below.
For details on individual calibrations, please read:
•
The separate IFU: Calibration
Interval
Calibration
Once per day
Gas calibration
Once per day, as well as after opening a new
carton of flow sensors
Volume calibration
Once per week
Flow linearity test
After Maintenance / Cleaning
Volume calibration
7.6
Using Ergostik Blueflow / Performing Measurements
Descriptions of the individual examinations can be found in the
separate IFU:
•
CPET
And for general use of the software:
•
BLUE CHERRY®