Section 6: System Specifications 

Power Requirements: 

72-84 W, 12-14 VDC input at 7.5 Amps

Nominal Operating Current: 

6 Amps DC

Over Current Protection: 

15 Amps

Maximum Operating Vacuum: 

27 in Hg

Ambient Operational Temperature: 

50°F-104°F (10°C-40°C)

Dimensions: 

6.70" H x 4.80" x 11.40" W
(17.02cm H x 12.92cm D x 28.96cm W)

Weight: 

11.4lbs (5.17kg)

Case: (optional) 

21.5” L x 15” H x 7.5” W
(54.6 cm L x 38.1cm H x 19.05cm W)

Case Weight: 

7.9 lbs. (3.6 kg) total

Free Air CFM @ 10 Vac in-Hg 

   0.38 

Free Air CFM @ 20 Vac in-Hg 

   0.16 

Features: 

• Approximately 45 to 60 minute operation with 8Ah lead acid battery.
• Reverse polarity protection.
• Optional AC cord with adapter for use with world mains connection.

Envirotecnics 

Tel: +34 872 080 542 

 

envirotecnics@envirote

cnics.com

www.envirotecnics.com

Summary of Contents for High Vacuum Sampler

Page 1: ...Rev 05 06 13 Part 12151001 Geotech High Vacuum Sampler Installation and Operation Manual...

Page 2: ...TEM OPERATION 7 SECTION 4 SYSTEM MAINTENANCE 8 SECTION 5 SYSTEM TROUBLESHOOTING 9 SECTION 6 SYSTEM SPECIFICATIONS 11 SECTION 7 SYSTEM SCHEMATIC 12 SECTION 8 REPLACEMENT PARTS AND ACCESSORIES 13 THE WA...

Page 3: ...d the WARNING message WARNING CAUTION A raised hand icon indicates CAUTION information that relates to a situation or condition that could lead to equipment malfunction or damage You should not procee...

Page 4: ...r Soil Vapor Extraction SVE systems Figure 1 1 is an example of the High Vacuum Sampler Each High Vacuum Sampler comes with a standard DC power cord Figure 1 2 to allow the unit to operate from a 12VD...

Page 5: ...4 Figure 1 2 Figure 1 3 Example of ALTEF 1 Liter Bag Envirotecnics Tel 34 872 080 542 envirotecnics envirotecnics com www envirotecnics com...

Page 6: ...5 Figure 1 4 Example of Tedlar 1 and 3 Liter Bags Envirotecnics Tel 34 872 080 542 envirotecnics envirotecnics com www envirotecnics com...

Page 7: ...fly and verify its operation Place your thumb over both ports Air should be blowing out the AIR OUT port and vacuum should be present at the VACUUM IN port Use the System Troubleshooting procedures in...

Page 8: ...the High Vacuum Sampler 2 Turn on the unit and allow the air flow to run prior to attaching your sample bag so that a good sample can be acquired 3 Push the valve stem in on the bag to open the valve...

Page 9: ...ap and water solution can be applied to a soft cloth and used to clean the exterior surfaces of the High Vacuum Sampler Do not immerse soak or directly spray liquids on the High Vacuum Sampler Equipme...

Page 10: ...ve The High Vacuum Sampler is protected from damage due to reverse polarity connection If ambient temperatures are in excess of 104 F 40 C then disconnect from power and let sit in a cool location Do...

Page 11: ...be sent back to a factory authorized repair location Problem Compressor turns on and there is vacuum along with air flowing out but there is no vacuum indicated on the gauge Solutions The pressure gau...

Page 12: ...W 17 02cm H x 12 92cm D x 28 96cm W Weight 11 4lbs 5 17kg Case optional 21 5 L x 15 H x 7 5 W 54 6 cm L x 38 1cm H x 19 05cm W Case Weight 7 9 lbs 3 6 kg total Free Air CFM 10 Vac in Hg 0 38 Free Air...

Page 13: ...12 Section 7 System Schematic Figure 7 1 High Vacuum Sampler Wiring Diagram Envirotecnics Tel 34 872 080 542 envirotecnics envirotecnics com www envirotecnics com...

Page 14: ...NCLUDES 10FT TUBING 57500008 ASSY POWER CORD DC W AMP 17500035 ADAPTER CIGARETTE TO CLIPS 77250000 BATTERY CHARGER MODULAR 12VDC 8Ah MALE CIG PLUG 51150127 CASE GEOCONTROL PRO CE 12150020 TEDLAR BAG 3...

Page 15: ...F DESCRIPTION REV DATE Previous Release 12 17 12 1570 Edited description for spare part 87050509 Added Revision History Table SP 05 06 13 Envirotecnics Tel 34 872 080 542 envirotecnics envirotecnics c...

Page 16: ...or to return of any equipment to our facilities please call our 800 number for appropriate location An RMA will be issued upon receipt of your request to return equipment which should include reasons...

Page 17: ...7958 FAX 303 322 7242 Envirotecnics Global Service SL Oficinas y Fabricaci n Calle dels Pirineus s n esquina Espinau 17460 Celr Girona Tel 872 080 542 Fax 872 080 543 Delegaci n Madrid Avda de Castill...

Reviews: