9
Example Flow Rates
Table 3-1
: Flow Rates
with 25’ (7.6m) Airline
@ 25 ft. (7.6 m) airline*
(3 ft. (0.9 m) submergence)
Pump Size:
Discharge Tube
Size:
Flow:
Approx. Geocontrol PRO Settings
Fill Time:
Discharge Time:
1.66 x 36"
(4 x 91 cm)
1/4" ID x 3/8" OD
(64 x 127 mm)
43 oz/min
(1.3 L/min)
10 sec.
7 sec.
1.66 x 18"
(4 x 46 cm)
33 oz/min
(1 L/min)
5 sec.
5 sec.
0.85 x 18"
(2.2 x 46 cm)
.17" ID x 1/4" OD
(43 x 64 mm)
20 oz/min
(0.6 L/min)
1 sec.
2 sec.
0.675 x 18"
(1.7 x 46 cm)
10 oz/min
(0.3 L/min)
1 sec.
2 sec.
Table 3-2
: Flow Rates with 150’ (46m) Airline
@ 150 ft. (46 m) airline*
(3 ft. (0.9 m) submergence)
Pump Size:
Discharge Tube
Size:
Flow:
Approx. Geocontrol PRO Settings
Fill Time:
Discharge Time:
1.66 x 36"
(4 x 91 cm)
1/4" ID x 3/8" OD
(64 x 127 mm)
9 oz/min
(260
mL/min)
20 sec.
55 sec.
1.66 x 18"
(4 x 46 cm)
6 oz/min
(175
mL/min)
15 sec.
40 sec.
0.85 x 18"
(2.2 x 46 cm)
.17" ID x 1/4" OD
(43 x 64 mm)
2 oz/min
(59
mL/min)
8 sec.
30 sec.
0.675 x 18"
(1.7 x 46 cm)
1 oz/min
(27
mL/min)
5 sec.
25 sec.
* Airline tubing size: .17”ID x ¼” OD (4 x 6 mm)
Speak with a Geotech representative to determine the best configuration to fulfill your
sampling needs.
Flow rates are based on 3’ (0.9 m) of pump submergence. Field
environments typically provide over 10’ (3 m) of submergence, which
increases flow.
Summary of Contents for 1.66x18
Page 1: ...Rev 4 15 2020 Part 11150324 PFAS Free Portable Bladder Pump Installation and Operation Manual...
Page 2: ......
Page 13: ...11 Figure 3 1 Tank Volume vs BP Consumption...
Page 26: ...24 Section 7 Replacement Parts List 1 66 Bladder Pump Components 18 model shown...
Page 28: ...26 850 Bladder Pump Components...
Page 30: ...28 675 Bladder Pump Components...
Page 32: ...30 Notes...