3.3.3 Current output
The current output of the 2550 is set via three
settings.
Low current
specifies the flow rate
represented by the minimum current signal
(4 or 0 mA).
High current
is the flow rate repre-
sented by the maximum current (20 mA), and
current
selects a 4 to 20 or a 0 to 20 mA span.
To change the current selections, first enter the
access code (section 3.2.1) then press the down
arrow until the display shows: