Geokon 4580 Instruction Manual Download Page 24

 

18 

APPENDIX C. USE OF SECOND ORDER POLYNOMIAL TO IMPROVE 

ACCURACY 

 

Most calibrations use linear coefficient to calculate pressure, loads strains etc. from measured 

readout “digits”. If the output of the sensor is not truly linear then there will be some inaccuracy 

introduced into the calculated value. Thus, even though the calibration accuracy may be 0.1% FS 

(because this is the accuracy of the calibration apparatus) yet the linearity may only be 0.5% F.S. 

so that the calculated value may differ from the true value by this larger amount. 

 

To recapture the 0.1% FS accuracy inherent in the calibration data it makes sense to use a second 

order polynomial to fit the calculated values to the actual calibration curve. The second order 

polynomial has the following characteristics. Calculated value = AR

2

 + BR + C where R is the 

observed reading on the readout box and A, B, C, are the three coefficients shown on the 

calibration report. Note that, unlike the linear method, this equation does not contain the item  

(R

1

 – R

0

) and thus requires a different treatment for establishing zero conditions. The coefficients 

shown on the calibration report are those developed under the conditions of temperature and 

barometric pressure experienced at the time of the calibration and which are shown on the 

calibration report. Theoretically, these could be used as is, but it is always good practice to 

establish and use zero conditions at the site since the zero may have shifted slightly due to rough 

handling during shipment and/or installation; and hence a slight adjustment of the “C” 

coefficient may be called for. Therefor it is recommended that the value of C be determined at 

the site for the conditions of temperature and barometric pressure experienced at that time.  

 

For example, using the data from the sample calibration report shown in Appendix D: 

At the site, at a temperature of T

0

 = 30

°

C and a barometric pressure of 1000 mbar, the observed 

zero pressure reading on a GK-404 readout (channel B) is R

=7420. Using the polynomial 

coefficient to calculate a new value of C.  

 

(7420)

2

 x 3.303E-09 + (7420) x 0. C = O 

 gives C = -13.634  

 

Note that the transducer shown is of a vented type, and hence no barometric correction is 

required. However, it is essential for vented types, that the moisture trap seal screw be removed 

before taking the zero reading (or any reading) so that the vent line is open to the atmosphere. 

 

 

 

Summary of Contents for 4580

Page 1: ... by any means without the written consent of Geokon The information contained herein is believed to be accurate and reliable However Geokon assumes no responsibility for errors omissions or misinterpretation The information herein is subject to change without notification Copyright 1995 2019 by Geokon Doc Rev Q 04 30 19 ...

Page 2: ......

Page 3: ...nnel There are no warranties except as stated herein There are no other warranties expressed or implied including but not limited to the implied warranties of merchantability and of fitness for a particular purpose Geokon is not responsible for any damages or losses caused to other equipment whether direct indirect incidental special or consequential which the purchaser may experience as a result ...

Page 4: ......

Page 5: ...he GK 405 7 3 3 GK 403 READOUT BOX OBSOLETE MODEL 8 3 3 1 Connecting Sensors with a 10 pin Bulkhead 8 3 3 2 Connecting Sensors with Bare Leads 8 3 3 3 Operating the GK 403 8 3 4 MEASURING TEMPERATURES 8 4 DATA REDUCTION 9 4 1 PRESSURE CALCULATION 9 4 2 TEMPERATURE CORRECTION 11 4 3 BAROMETRIC CORRECTION 11 4 3 1 Vented Transducers 13 4 4 ENVIRONMENTAL FACTORS 13 5 MAINTENANCE 14 6 TROUBLESHOOTING ...

Page 6: ...ADINGS 7 FIGURE 7 SAMPLE MODEL 4580 CALIBRATION REPORT 19 TABLES TABLE 1 ENGINEERING UNITS MULTIPLICATION FACTORS 10 TABLE 2 MODEL 4580 PRESSURE TRANSDUCER SPECIFICATIONS 16 TABLE 3 THERMISTOR RESISTANCE VERSUS TEMPERATURE 17 EQUATIONS EQUATION 1 DIGITS CALCULATION 9 EQUATION 2 CONVERT DIGITS TO PRESSURE 9 EQUATION 3 TEMPERATURE CORRECTION 11 EQUATION 4 BAROMETRIC CORRECTION 11 EQUATION 5 CORRECTE...

Page 7: ...ic effects need to be eliminated such as in the weir and stream level measurements In the case where the sensor is used as a barometer the capsule is evacuated The vibrating wire is attached to the capsule on one end and to the rigid housing on the other External pressure applied to the capsule causes it to compress and exert a force on the vibrating wire gauge and this produces a change in wire t...

Page 8: ...etween the sensor being vertically up and vertically down is normal This phenomenon will not be a problem if the sensor remains stationary or is in the same orientation during use Additionally these sensors are sensitive to shock particularly the very low pressure versions A shift in the zero reading can occur during shipping or rough handling Normally this will not affect the operation of the sen...

Page 9: ...th the transducers held under water to prevent air from reentering the sensor Air trapped in the chamber can cause inaccurate and or unstable pressure readings in changing temperature environments The vented versions incorporate a moisture trap on the end of the vent line Desiccant inside the moisture trap prevents moisture from entering the inside of the sensor capsule The vent trap screw must be...

Page 10: ...r these types of installations it is recommended that the transducer be supplied with enough cable to reach the installation depth plus extra cable to pass through drilling equipment rods casing etc Cable used for making splices should be a high quality twisted pair type with 100 shielding and an integral shield drain wire When splicing it is very important that the shield drain wires be spliced t...

Page 11: ...nal boxes available from Geokon can be ordered with lightning protection built in The terminal board used to make the gauge connections has provision for the installation of plasma surge arrestors Lightning Arrestor Boards LAB 3 can also be incorporated into the terminal box The terminal box must be connected to an earth ground for these levels of protection to be effective If the instruments will...

Page 12: ...rs of the sensor conductors with blue representing the shield bare To turn the GK 404 on press the ON OFF button on the front panel of the unit The initial startup screen will be displayed After approximately one second the GK 404 will start taking readings and display them based on the settings of the POS and MODE buttons The unit display from left to right is as follows The current Position Set ...

Page 13: ...rs with Bare Leads Attach the GK 403 2 flying leads to the bare leads of a Geokon vibrating wire sensor by connecting each of the clips on the leads to the matching colors of the sensor conductors with blue representing the shield bare 3 2 3 Operating the GK 405 Press the button labeled POWER ON A blue light will begin blinking signifying that the Remote Module is waiting to connect to the handhel...

Page 14: ...yed above the gauge reading in degrees centigrade 5 Press the Store button to record the value displayed If the no reading displays or the reading is unstable see Section 6 for troubleshooting suggestions The unit will automatically turn off after approximately two minutes to conserve power Consult the GK 403 Instruction Manual for additional information 3 4 Measuring Temperatures All vibrating wi...

Page 15: ...ssure To convert digits to pressure using a linear calibration coefficient the following equation applies See Appendix C for use of second order polynomial Puncorrected R1 R0 G F1 Equation 2 Convert Digits to Pressure Where R0 represents the initial reading taken at installation usually the zero reading R1 represents the current reading G is the calibration factor required to convert from digits t...

Page 16: ...000133 101320 0001 1 001 1 Table 1 Engineering Units Multiplication Factors Note Due to changes in specific gravity with temperature the factors for mercury and water in the above table are approximate In the special case where the Model 4580 is being used as a Barometer it is necessary to add the atmospheric pressure at the time of the initial reading This pressure would best be obtained locally ...

Page 17: ...tial temperature K is the thermal coefficient from the supplied calibration report F1 is an optional multiplier to convert to engineering units see Table 1 The calculated correction would then be added to the Pressure calculated using Equation 2 If the engineering units were converted remember to apply the same conversion to the calculated temperature correction 4 3 Barometric Correction Since the...

Page 18: ...ion is to use pressure transducers that are vented to the atmosphere as described in Section 4 3 1 With vented transducers barometric pressure finds its way to both the inside and the outside of the pressure capsule and is thus automatically canceled Equation 5 describes the pressure calculation with both temperature and barometric corrections applied Pcorrected R1 R0 G T1 T0 K S1 S0 F2 Equation 5...

Page 19: ...desiccant capsules are blue when fresh they will gradually turn pink as they absorb moisture When they have turned light pink in color they should be replaced Contact Geokon for replacement capsules 4 4 Environmental Factors Since the purpose of the transducer installation is to monitor site conditions factors that can affect these conditions should always be observed and recorded Seemingly minor ...

Page 20: ...e sensor is installed in water with high concentrations of silts etc it may require cleaning from time to time This can be accomplished by gently flushing water through the transducer first through the P 1 port and then back through the seal screw hole Repeat this several times until the water passing through is clear Remember not to apply excess pressure to the sensor ...

Page 21: ...adings automatically are the swept frequency excitation settings correct Try reading the transducer on a different readout position For instance channel A of the GK 404 and GK 405 might be able to read the sensor To convert the Channel A period display to digits use Equation 1 in Section 4 1 Is there a source of electrical noise nearby Most probable sources of electrical noise are motors generator...

Page 22: ...ength 110 mm 4 33 172 mm 6 75 172 mm 6 75 112 mm 4 5 Weight 1 3 kg 2 8 lbs 1 5 kg 3 3 lbs 2 0 kg 4 4 lbs 4 0 kg 9 0 lbs Table 2 Model 4580 Pressure Transducer Specifications Notes Other ranges available upon request The accuracy of the calibration equipment 0 1 FS linearity available upon request Also 0 1 accuracy can be achieved by using a second order polynomial instead of a linear calibration f...

Page 23: ... 2 83 107 9 123 77 66K 36 8006 4 1363 44 331 5 84 105 2 124 72 81K 35 7618 5 1310 45 321 2 85 102 5 125 68 30K 34 7252 6 1260 46 311 3 86 99 9 126 64 09K 33 6905 7 1212 47 301 7 87 97 3 127 60 17K 32 6576 8 1167 48 292 4 88 94 9 128 56 51K 31 6265 9 1123 49 283 5 89 92 5 129 53 10K 30 5971 10 1081 50 274 9 90 90 2 130 49 91K 29 5692 11 1040 51 266 6 91 87 9 131 46 94K 28 5427 12 1002 52 258 6 92 8...

Page 24: ...nts shown on the calibration report are those developed under the conditions of temperature and barometric pressure experienced at the time of the calibration and which are shown on the calibration report Theoretically these could be used as is but it is always good practice to establish and use zero conditions at the site since the zero may have shifted slightly due to rough handling during shipm...

Page 25: ...19 APPENDIX D TYPICAL CALIBRATION REPORT Figure 7 Sample Model 4580 Calibration Report ...

Reviews: