4.3.1 Vented Transducers
Vented transducers are designed to eliminate barometric effects. The space inside the
transducer is not hermetically sealed and evacuated (see Figure 1) but is connected via a
tube (integral with the cable) to the atmosphere. A chamber containing desiccant capsules
is attached to the end of the tube to prevent moisture from entered the transducer cavity.
As supplied, the outer end of the desiccant chamber is closed by means of a seal screw to
keep the desiccant fresh during storage and transportation.
THE SEAL SCREW MUST
BE REMOVED BEFORE THE TRANSDUCER IS PUT INTO SERVICE!
The
desiccant capsules are blue when fresh, they will gradually turn pink as they absorb
moisture. When they have turned light pink in color they should be replaced. Contact
Geokon for replacement capsules.
4.4 Environmental Factors
Since the purpose of the transducer installation is to monitor site conditions, factors that can
affect these conditions should always be observed and recorded. Seemingly minor effects can
have a real influence on the behavior of the structure being monitored and can give an early
indication of potential problems. Some of the factors include, but are not limited to; blasting,
rainfall, tidal levels, excavation and fill levels and sequences, traffic, temperature and barometric
changes (and other weather conditions), changes in personnel, nearby construction activities,
seasonal changes, etc.