![Geokon 4500MLP Instruction Manual Download Page 27](http://html.mh-extra.com/html/geokon/4500mlp/4500mlp_instruction-manual_2214562027.webp)
21
From
→
To
↓
psi
"H2O
'H2O
mm H20
m H20
"HG
mm HG
atm
mbar
bar
kPa
MPa
psi
1
.036127 .43275 .0014223 1.4223
.49116 .019337 14.696 .014503 14.5039 .14503
145.03
"H2O
27.730
1
12
.039372
39.372
13.596
.53525
406.78
.40147
401.47
4.0147
4016.1
'H2O
2.3108
.08333
1
.003281
3.281
1.133
.044604 33.8983 .033456 33.4558
.3346
334.6
mm H20
704.32
25.399 304.788
1
1000
345.32
13.595
10332
10.197
10197
101.97
101970
m H20
.70432 .025399 .304788
.001
1
.34532 .013595 10.332 .010197 10.197
.10197
101.97
"HG
2.036
.073552 .882624 .0028959 2.8959
1
.03937
29.920 .029529 29.529
.2953
295.3
mm HG
51.706
1.8683 22.4196 .073558
73.558
25.4
1
760
.75008
750.08
7.5008
7500.8
atm
.06805 .002458 .029499 .0000968
.0968
.03342 .001315
1
.000986 .98692 .009869
9.869
mbar
68.947
2.4908 29.8896 .098068
98.068
33.863
1.3332
1013.2
1
1000
10
10000
bar
.068947 .002490 .029889 .0000981 .098068 .033863 .001333 1.0132
.001
1
.01
10
kPa
6.8947
.24908 2.98896 .0098068 9.8068
3.3863
.13332 101.320
.1
100
1
1000
MPa
.006895 .000249 .002988 .0000098 .009807 .003386 .000133 .101320
.0001
.1
.001
1
Table 2 - Engineering Units Multiplication Factors
(Note: Due to changes in specific gravity with temperature, the factors for mercury and water in
the above table are approximations.)
5.2 Temperature Correction
The materials used in the construction of Geokon’s vibrating wire piezometers have been
carefully selected to minimize thermal effects; however, most units still have a slight temperature
coefficient. Consult the calibration sheet supplied with the instrument to obtain the coefficient
for the individual piezometer.
Since piezometers are normally installed in a tranquil and constant temperature environment,
corrections are normally not required. If this is not the case for the selected installation,
corrections can be made using the internal thermistor for temperature measurement. See Section
4.4 for instructions regarding obtaining the piezometer temperature.
The temperature correction equation is as follows:
Temperature Correction
=
(Current Temperature - Initial Temperature)
×
Thermal Factor
Or
P
T
= (T
1
-T
0
) x K
Equation 3 - Temperature Correction
The calculated correction would then be added to the pressure calculated using Equation 2. If the
engineering units were converted, remember to apply
the same conversion
to the calculated
temperature correction.
For example: If the initial temperature was 22
°
C, and the current temperature is 15
°
C, and the
thermal factor (K on the calibration report,) is +0.1319 kPa per
°
C rise. The temperature
correction is +0.1319(15-22) = -0.92 kPa. Refer to the calibration report provided with the
instrument for the initial temperature and thermal factor.